Performance of a New Leadless Implantable Cardiac Monitor in Detecting and Quantifying Atrial Fibrillation Results of the XPECT Trial
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00023001%3A_____%2F10%3A00055981" target="_blank" >RIV/00023001:_____/10:00055981 - isvavai.cz</a>
Result on the web
<a href="http://circep.ahajournals.org/content/3/2/141.full.pdf+html" target="_blank" >http://circep.ahajournals.org/content/3/2/141.full.pdf+html</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1161/CIRCEP.109.877852" target="_blank" >10.1161/CIRCEP.109.877852</a>
Alternative languages
Result language
angličtina
Original language name
Performance of a New Leadless Implantable Cardiac Monitor in Detecting and Quantifying Atrial Fibrillation Results of the XPECT Trial
Original language description
Background-Current methods for detecting atrial fibrillation (AF) have limited diagnostic yield. Continuous monitoring with automatic arrhythmia detection and classification may improve detection of symptomatic and asymptomatic AF and subsequent patienttreatment. The study purpose was to quantify the performance of the first implantable leadless cardiac monitor (ICM) with dedicated AF detection capabilities. Methods and Results-Patients (n247) with an implanted ICM (Reveal XT, Medtronic Inc, Minneapolis, Minn) who were likely to present with paroxysmal AF were selected. A special Holter device stored 46 hours of subcutaneously recorded ECG, ICM markers, and 2 surface ECG leads. The ICM automatic arrhythmia classification was compared with the core laboratory classification of the surface ECG. Of the 206 analyzable Holter recordings collected, 76 (37%) contained at least 1 episode of core laboratory classified AF. The sensitivity, specificity, positive predictive value, and negative pr
Czech name
—
Czech description
—
Classification
Type
J<sub>x</sub> - Unclassified - Peer-reviewed scientific article (Jimp, Jsc and Jost)
CEP classification
FA - Cardiovascular diseases including cardio-surgery
OECD FORD branch
—
Result continuities
Project
—
Continuities
N - Vyzkumna aktivita podporovana z neverejnych zdroju
Others
Publication year
2010
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Name of the periodical
Circulation-Arrhythmia and Electrophysiology
ISSN
1941-3149
e-ISSN
—
Volume of the periodical
3
Issue of the periodical within the volume
2
Country of publishing house
US - UNITED STATES
Number of pages
7
Pages from-to
141-147
UT code for WoS article
000276893200006
EID of the result in the Scopus database
—