Five-year outcomes of patients supported with HeartMate 3: a single-centre experience
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00023001%3A_____%2F21%3A00081150" target="_blank" >RIV/00023001:_____/21:00081150 - isvavai.cz</a>
Result on the web
<a href="https://academic.oup.com/ejcts/article-abstract/59/6/1155/6135342?redirectedFrom=fulltext" target="_blank" >https://academic.oup.com/ejcts/article-abstract/59/6/1155/6135342?redirectedFrom=fulltext</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1093/ejcts/ezab018" target="_blank" >10.1093/ejcts/ezab018</a>
Alternative languages
Result language
angličtina
Original language name
Five-year outcomes of patients supported with HeartMate 3: a single-centre experience
Original language description
OBJECTIVES: The HeartMate 3 left ventricular assist device was first implanted in 2014 and received the Conformite Europeenne mark in 2015. Since then, several trials demonstrated its high haemocompatibility associated with good survival and low adverse events rates. Herein, we report our institutional experience with patients supported with HeartMate 3 for 5 years. METHODS: This prospective cohort study included patients receiving a HeartMate 3 implantation in 2014 as part of the HeartMate 3 Conformite Europeenne Mark clinical trial. Patients had follow-up visits every 3 months while on left ventricular assist device support, and all patients completed the 5-year follow-up. The primary end point was survival at 5 years. Secondary end points included adverse events, health status and quality of life. RESULTS: Eight patients (men: 75%) aged 59 years (min-max: 52-66 years) were enrolled. At 5 years, survival was 100%. Patients remained on support for a median time of 1825 days (min-max: 101-1825 days); 2 patients successfully received cardiac transplants. No right heart failure, haemolysis, pump thrombosis, pump malfunction or neurological events occurred in any patients. A driveline infection was observed in 6 patients (0.25 events/patient-year). Compared to baseline, a significant improvement in quality of life and in New York Heart Association functional class was noted after the implant and for the whole follow-up time. A slight decline in kidney function and in the 6-min walk test results occurred after 3 years. CONCLUSIONS: This study reports the longest single-centre follow-up of the HeartMate 3, showing excellent haemocompatibility over time with high survival and low complication rates at 5 years.
Czech name
—
Czech description
—
Classification
Type
J<sub>imp</sub> - Article in a specialist periodical, which is included in the Web of Science database
CEP classification
—
OECD FORD branch
30201 - Cardiac and Cardiovascular systems
Result continuities
Project
—
Continuities
N - Vyzkumna aktivita podporovana z neverejnych zdroju
Others
Publication year
2021
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Name of the periodical
European journal of cardio-thoracic surgery
ISSN
1010-7940
e-ISSN
—
Volume of the periodical
59
Issue of the periodical within the volume
6
Country of publishing house
GB - UNITED KINGDOM
Number of pages
9
Pages from-to
1155-1163