Regenerated Bone Quality as a Determinant of Bone Turnover and Prognosis in Short Plateau Implants: A Finite Element Study
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00064165%3A_____%2F25%3A10504957" target="_blank" >RIV/00064165:_____/25:10504957 - isvavai.cz</a>
Result on the web
<a href="https://verso.is.cuni.cz/pub/verso.fpl?fname=obd_publikace_handle&handle=HPaBgPzm4P" target="_blank" >https://verso.is.cuni.cz/pub/verso.fpl?fname=obd_publikace_handle&handle=HPaBgPzm4P</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.3390/prosthesis7050123" target="_blank" >10.3390/prosthesis7050123</a>
Alternative languages
Result language
angličtina
Original language name
Regenerated Bone Quality as a Determinant of Bone Turnover and Prognosis in Short Plateau Implants: A Finite Element Study
Original language description
Background/Objectives: Finite element analysis (FEA) can predict biomechanical performance of dental implants in compromised bone. In the posterior maxilla, low bone density, thin cortex, and variable regenerated bone stiffness may lead to pathological peri-implant strains. This study examined the effects of implant diameter, cortical thickness, cancellous bone type, and regenerated bone elasticity on strain distribution in short plateau (Bicon SHORT (R)) implants. Methods: Three-dimensional FEA models of type III and IV maxillae with cortical layers of 1.0, 0.75, and 0.5 mm were developed. Implants of 4.5, 5.0, and 6.0 mm diameter were tested, with regenerated bone elasticity set to 25-100% of cortical values. An oblique load of 120.9 N at 75 degrees was applied under full osseointegration, and first principal strains were compared with Frost's 3000 mu epsilon threshold. Results: Cortical strains remained at physiological levels, but cancellous bone in type IV often exceeded 3000 mu epsilon, especially with smaller diameters and low regenerated stiffness. Enlarging implant diameter to 6.0 mm lowered cancellous maximal first principal strain by up to 56% in type III and 36% in type IV bone. Reduced regenerated bone elasticity markedly increased risk, particularly with cortical thickness < 0.75 mm. Conclusions: Biomechanical risk depends on implant diameter and regenerated bone quality. Wide short implants (6.0 mm) most effectively limited pathological strain under low cortical support and poor regenerated stiffness. Patient-specific FEA may guide implant choice and improve outcomes in atrophic maxilla rehabilitation.
Czech name
—
Czech description
—
Classification
Type
J<sub>imp</sub> - Article in a specialist periodical, which is included in the Web of Science database
CEP classification
—
OECD FORD branch
30208 - Dentistry, oral surgery and medicine
Result continuities
Project
—
Continuities
V - Vyzkumna aktivita podporovana z jinych verejnych zdroju
Others
Publication year
2025
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Name of the periodical
Prosthesis
ISSN
2673-1592
e-ISSN
2673-1592
Volume of the periodical
7
Issue of the periodical within the volume
5
Country of publishing house
CH - SWITZERLAND
Number of pages
16
Pages from-to
123