Inter-comparison of phantoms for CT numbers to relative electron density (RED)/physical density calibration and influence to dose calculation in TPS

The result's identifiers

  • Result code in IS VaVaI

    <a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00064203%3A_____%2F19%3A10395254" target="_blank" >RIV/00064203:_____/19:10395254 - isvavai.cz</a>

  • Alternative codes found

    RIV/86652052:_____/19:N0000003

  • Result on the web

    <a href="https://doi.org/10.1088/1742-6596/1248/1/012046" target="_blank" >https://doi.org/10.1088/1742-6596/1248/1/012046</a>

  • DOI - Digital Object Identifier

    <a href="http://dx.doi.org/10.1088/1742-6596/1248/1/012046" target="_blank" >10.1088/1742-6596/1248/1/012046</a>

Alternative languages

  • Result language

    angličtina

  • Original language name

    Inter-comparison of phantoms for CT numbers to relative electron density (RED)/physical density calibration and influence to dose calculation in TPS

  • Original language description

    Six different phantoms (2x CIRS, 1x Agmeco, 1x QUASAR, 1x Gammex, 1x CatPhan) and Varian formulas were used to evaluate their suitability for CT numbers to relative electron density (RED) calibration in the treatment planning system at two centres using two CT scanners. The variability in calibration curve and the influence to dose calculation in the TPS were estimated for 6 MV and 18 MV. The presence of scattering material around inhomogeneities and influence of CT spectra (90 kV, 120 kV, 140 kV) were evaluated. All phantoms were found convenient for calibration except CatPhan where large deviations were observed for dense bone. Phantoms without scatter material around inhomogeneities should be used with caution. Varian formula slightly overestimates calibration curve in the region of 200 - 300 HU. The tolerance level of 10% suggested for RED during commissioning is reasonable as it can cause change in dose of 3% and 1.5% for 6 MV and 18 MV respectively as a very conservative estimate. Calibration curves should be acquired for all scanning protocols with different CT spectra.

  • Czech name

  • Czech description

Classification

  • Type

    D - Article in proceedings

  • CEP classification

  • OECD FORD branch

    30204 - Oncology

Result continuities

  • Project

  • Continuities

    I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace

Others

  • Publication year

    2019

  • Confidentiality

    S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů

Data specific for result type

  • Article name in the collection

    Journal of Physics: Conference Series

  • ISBN

  • ISSN

    1742-6588

  • e-ISSN

  • Number of pages

    6

  • Pages from-to

  • Publisher name

    Institute of Physics Publishing

  • Place of publication

    Bristol

  • Event location

    Kuala Lumpur

  • Event date

    Nov 11, 2018

  • Type of event by nationality

    WRD - Celosvětová akce

  • UT code for WoS article

    000492968100046