Comparison of toric intraocular lens tilt and decentration measurement using dynamic Purkinje-meter and anterior segment optical coherence tomography
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00064203%3A_____%2F25%3A10481260" target="_blank" >RIV/00064203:_____/25:10481260 - isvavai.cz</a>
Alternative codes found
RIV/68407700:21110/25:00376727 RIV/68407700:21460/25:00376727 RIV/00216208:11130/25:10481260
Result on the web
<a href="https://verso.is.cuni.cz/pub/verso.fpl?fname=obd_publikace_handle&handle=rCPn00zY5_" target="_blank" >https://verso.is.cuni.cz/pub/verso.fpl?fname=obd_publikace_handle&handle=rCPn00zY5_</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.5507/bp.2024.017" target="_blank" >10.5507/bp.2024.017</a>
Alternative languages
Result language
angličtina
Original language name
Comparison of toric intraocular lens tilt and decentration measurement using dynamic Purkinje-meter and anterior segment optical coherence tomography
Original language description
AIMS: To present a new method of dynamic Purkinje-metry and to verify it by comparison with a commercially available anterior segment optical coherence tomography CASIA2. PATIENTS AND METHODS: A dynamic Purkinje-meter with a movable fixation target was assembled. A coaxial circular pattern formed by infrared LEDs was projected onto the eye and evoked Purkinje images (1st, 3rd, 4th = P1, P3, P4). The measurement was performed on 29 eyes with an implanted toric IOL (intraocular lens), under mydriatic conditions, with reference to the visual axis. The IOL tilt was calculated from the position of a fixation target at the moment of P3 and P4 superposition. The IOL decentration was determined based on the relative position of P1 during on-axis fixation and of P3 and P4 superposition during off-axis fixation. A custom-developed software was used for distance measurements. Using CASIA2, the IOL position was fully calculated by the device. RESULTS: The mean absolute difference between CASIA2 and Purkinje-meter values was 0.6° +- 0.4° for the tilt magnitude and 10° +- 10° for the tilt direction, and 0.11 mm +- 0.08 mm for the decentration magnitude and 16° +- 14° for the decentration direction. There was no statistically significant difference between the values determined by the two methods for the tilt and decentration direction. The differences were statistically significant for the tilt and decentration magnitude. CONCLUSION: The values of IOL tilt and decentration direction are similar for both devices. The values of IOL tilt and decentration magnitude measured by Purkinje-meter are higher than those from CASIA2, but overall, they correspond to the values presented in other published studies.
Czech name
—
Czech description
—
Classification
Type
J<sub>imp</sub> - Article in a specialist periodical, which is included in the Web of Science database
CEP classification
—
OECD FORD branch
30207 - Ophthalmology
Result continuities
Project
—
Continuities
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Others
Publication year
2025
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Name of the periodical
Biomedical Papers
ISSN
1213-8118
e-ISSN
1804-7521
Volume of the periodical
169
Issue of the periodical within the volume
1
Country of publishing house
CZ - CZECH REPUBLIC
Number of pages
10
Pages from-to
56-65
UT code for WoS article
001245524500001
EID of the result in the Scopus database
2-s2.0-86000728551