Experimental and MC determination of HPGe detector efficiency in the 40-2754 keV energy range for measuring point source geometry with the source-to-detector distance of 25 cm
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00177016%3A_____%2F06%3A%230000129" target="_blank" >RIV/00177016:_____/06:#0000129 - isvavai.cz</a>
Result on the web
—
DOI - Digital Object Identifier
—
Alternative languages
Result language
angličtina
Original language name
Experimental and MC determination of HPGe detector efficiency in the 40-2754 keV energy range for measuring point source geometry with the source-to-detector distance of 25 cm
Original language description
Monte Carlo model for HPGe detector was created and measurement efficiency calculated in the energy range 40-2754 keV. The model was validated using experimental values for a point source and the source-todetector distance of 25 cm.
Czech name
—
Czech description
—
Classification
Type
J<sub>x</sub> - Unclassified - Peer-reviewed scientific article (Jimp, Jsc and Jost)
CEP classification
BG - Nuclear, atomic and molecular physics, accelerators
OECD FORD branch
—
Result continuities
Project
—
Continuities
R - Projekt Ramcoveho programu EK
Others
Publication year
2006
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Name of the periodical
Applied Radiation and Isotopes
ISSN
0969-8043
e-ISSN
—
Volume of the periodical
64
Issue of the periodical within the volume
10-11
Country of publishing house
GB - UNITED KINGDOM
Number of pages
4
Pages from-to
—
UT code for WoS article
—
EID of the result in the Scopus database
—