Precision Inductance Measurement on High Precision Capacitance Bridge
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00177016%3A_____%2F08%3A%230000245" target="_blank" >RIV/00177016:_____/08:#0000245 - isvavai.cz</a>
Result on the web
—
DOI - Digital Object Identifier
—
Alternative languages
Result language
angličtina
Original language name
Precision Inductance Measurement on High Precision Capacitance Bridge
Original language description
For high precision impedance measurement only a very limited range of commercial instruments is available. The most accurate commercially available instruments for impedance measurements are capacitance bridges. Precise instruments for inductance or general impedance measurements are not available on the market. This paper describes novel methods for calibration of inductance reference standards from 1 ?H to 10 kH using an accurate capacitance bridge with uncertainty on 1 kHz better than 50 ppm for 100?H to 10 kH and better than 500 ppm for 1 ?H and 10 ?H. Main attention is paid to new measurement method using active or passive T-circuit for inductances from 1 ?H to 1 H.
Czech name
—
Czech description
—
Classification
Type
D - Article in proceedings
CEP classification
JB - Sensors, detecting elements, measurement and regulation
OECD FORD branch
—
Result continuities
Project
—
Continuities
R - Projekt Ramcoveho programu EK
Others
Publication year
2008
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Article name in the collection
Conference on Precision Electromagnetic Measurements Digest
ISBN
978-1-4244-2400-9
ISSN
—
e-ISSN
—
Number of pages
2
Pages from-to
—
Publisher name
IEEE, 345 E 47th St., New York, NY 10017 USA
Place of publication
Boulder, USA
Event location
Boulder, USA
Event date
Jan 1, 2008
Type of event by nationality
WRD - Celosvětová akce
UT code for WoS article
—