All

What are you looking for?

All
Projects
Results
Organizations

Quick search

  • Projects supported by TA ČR
  • Excellent projects
  • Projects with the highest public support
  • Current projects

Smart search

  • That is how I find a specific +word
  • That is how I leave the -word out of the results
  • “That is how I can find the whole phrase”

Calibration of Industrial Platinum Resistance Thermometers up to 700 °C

The result's identifiers

  • Result code in IS VaVaI

    <a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00177016%3A_____%2F13%3A%230001004" target="_blank" >RIV/00177016:_____/13:#0001004 - isvavai.cz</a>

  • Result on the web

  • DOI - Digital Object Identifier

Alternative languages

  • Result language

    angličtina

  • Original language name

    Calibration of Industrial Platinum Resistance Thermometers up to 700 °C

  • Original language description

    Industrial grade platinum resistance thermometers were calibrated in the temperature range 200 °C ? 700 °C. Both wire wound and thin film sensor based thermometers were investigated. The calibrations were performed in a vertical Cs-heat pipe furnace andin a horizontal Na heat pipe furnace. The reference thermometer was a SPRT calibrated at fixed points up to Al point. In addition to calibration, various thermal tests including immersion measurements and thermal cycling tests were performed. The stability of the sensors was determined by monitoring the ice point resistance. Possible contamination of the sensors were determined by measuring the resistance ratio R(30 °C)/R(10 °C) several times during the measurement period. The calibration curves were compared with ICE 60751 and ITS-90 reference functions. Considerable changes were found in all tested thermometers.. The wire wound sensors were more stable than the thin film sensors.

  • Czech name

  • Czech description

Classification

  • Type

    D - Article in proceedings

  • CEP classification

    BJ - Thermodynamics

  • OECD FORD branch

Result continuities

  • Project

    <a href="/en/project/7A11082" target="_blank" >7A11082: Metrology for improved power plant efficiency</a><br>

  • Continuities

    P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)

Others

  • Publication year

    2013

  • Confidentiality

    S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů

Data specific for result type

  • Article name in the collection

    Symposium on temperature and thermal measurements in industry and science

  • ISBN

    978-972-8574-15-4

  • ISSN

  • e-ISSN

  • Number of pages

    1

  • Pages from-to

    201

  • Publisher name

  • Place of publication

    Madeira

  • Event location

    Madeira

  • Event date

    Jan 1, 2013

  • Type of event by nationality

    WRD - Celosvětová akce

  • UT code for WoS article