Calibration of power sensors for low-power measurement
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00177016%3A_____%2F19%3AN0000082" target="_blank" >RIV/00177016:_____/19:N0000082 - isvavai.cz</a>
Result on the web
<a href="http://rfmw.cmi.cz/index.php/download/54-bpguides.html" target="_blank" >http://rfmw.cmi.cz/index.php/download/54-bpguides.html</a>
DOI - Digital Object Identifier
—
Alternative languages
Result language
angličtina
Original language name
Calibration of power sensors for low-power measurement
Original language description
This report describes the method for calibration of power sensors for low-power measurement. It focuses on the most common measurement configuration of three sensors (one leveling sensor, the standard or calibrated sensor and the sensor under test). As the first approach, it is assumed that a calibrated diode sensor is available at the laboratory. Later on, a correction is introduced which uses a padding attenuator with which the power level incident onto the diode sensor under test is reduced: by doing this, the power readings of the diode sensor under test and a calibrated mid-power thermocouple can be compared.
Czech name
—
Czech description
—
Classification
Type
V<sub>souhrn</sub> - Summary research report
CEP classification
—
OECD FORD branch
20201 - Electrical and electronic engineering
Result continuities
Project
<a href="/en/project/8B16005" target="_blank" >8B16005: Development of RF and microwave metrology capability</a><br>
Continuities
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Others
Publication year
2019
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Number of pages
18
Place of publication
Torrejon de Ardoz, Španělsko
Publisher/client name
EURAMET
Version
—