Interlaboratory comparison of battery impedance analyzers calibration

The result's identifiers

  • Result code in IS VaVaI

    <a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00177016%3A_____%2F23%3AN0000053" target="_blank" >RIV/00177016:_____/23:N0000053 - isvavai.cz</a>

  • Result on the web

    <a href="https://www.sciencedirect.com/science/article/pii/S0263224123007406" target="_blank" >https://www.sciencedirect.com/science/article/pii/S0263224123007406</a>

  • DOI - Digital Object Identifier

    <a href="http://dx.doi.org/10.1016/j.measurement.2023.113176" target="_blank" >10.1016/j.measurement.2023.113176</a>

Alternative languages

  • Result language

    angličtina

  • Original language name

    Interlaboratory comparison of battery impedance analyzers calibration

  • Original language description

    The paper reports a results of series of interlaboratory comparisons of low impedance measurements at frequencies relevant for electrochemical impedance spectroscopy (EIS) of commercial lithium-ion cells. Two comparisons are presented. The first, bilateral comparison has focused on low impedance standards calibration in a full complex plane using digital sampling setups. The second comparison has focused on calibration and use of commercial 4-terminal battery EIS meters. Both comparisons have covered the impedance range from 50 μΩ to 100 mΩ across the full complex plane in a frequency range from 0.01 Hz up to 5 kHz. Finally, the paper summarizes practices identified as critical for achieving measurement compatibility among various labs.

  • Czech name

  • Czech description

Classification

  • Type

    J<sub>SC</sub> - Article in a specialist periodical, which is included in the SCOPUS database

  • CEP classification

  • OECD FORD branch

    20201 - Electrical and electronic engineering

Result continuities

  • Project

    <a href="/en/project/8B18008" target="_blank" >8B18008: Quality assessment of electric vehicle Li-ion batteries for second use applications</a><br>

  • Continuities

    P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)<br>I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace

Others

  • Publication year

    2023

  • Confidentiality

    S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů

Data specific for result type

  • Name of the periodical

    Measurement

  • ISSN

    0263-2241

  • e-ISSN

  • Volume of the periodical

  • Issue of the periodical within the volume

    218

  • Country of publishing house

    IT - ITALY

  • Number of pages

    8

  • Pages from-to

    113176-113183

  • UT code for WoS article

  • EID of the result in the Scopus database

    2-s2.0-85163161321