Long-term stability of the Leksell Gamma Knife((R)) Perfexion (TM) patient positioning system (PPS)

The result's identifiers

  • Result code in IS VaVaI

    <a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216208%3A11110%2F14%3A10290740" target="_blank" >RIV/00216208:11110/14:10290740 - isvavai.cz</a>

  • Alternative codes found

    RIV/68407700:21340/14:00228872 RIV/00064203:_____/14:10294382

  • Result on the web

    <a href="http://dx.doi.org/10.1118/1.4866225" target="_blank" >http://dx.doi.org/10.1118/1.4866225</a>

  • DOI - Digital Object Identifier

    <a href="http://dx.doi.org/10.1118/1.4866225" target="_blank" >10.1118/1.4866225</a>

Alternative languages

  • Result language

    angličtina

  • Original language name

    Long-term stability of the Leksell Gamma Knife((R)) Perfexion (TM) patient positioning system (PPS)

  • Original language description

    Purpose: To assess the long-term mechanical stability and accuracy of the patient positioning system (PPS) of the Leksell Gamma Knife (R) Perfexion (TM) (LGK PFX). Methods: The mechanical stability of the PPS of the LGK PFX was evaluated using measurements obtained between September 2007 and June 2011. Three methods were employed to measure the deviation of the coincidence of the radiological focus point (RFP) and the PPS calibration center point (CCP). In the first method, the onsite diode test tool with single diode detector was used together with the 4 mm collimator on a daily basis. In the second method, a service diode test tool with three diode detectors was used biannually at the time of the routine preventive maintenance. The test performed with the service diode test tool measured the deviations for all three collimators 4, 8, and 16 mm and also for three different positions of the PPS. The third method employed the conventional film pin-prick method. This test was performed a

  • Czech name

  • Czech description

Classification

  • Type

    J<sub>x</sub> - Unclassified - Peer-reviewed scientific article (Jimp, Jsc and Jost)

  • CEP classification

    BO - Biophysics

  • OECD FORD branch

Result continuities

  • Project

  • Continuities

    V - Vyzkumna aktivita podporovana z jinych verejnych zdroju

Others

  • Publication year

    2014

  • Confidentiality

    S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů

Data specific for result type

  • Name of the periodical

    Medical Physics

  • ISSN

    0094-2405

  • e-ISSN

  • Volume of the periodical

    41

  • Issue of the periodical within the volume

    3

  • Country of publishing house

    US - UNITED STATES

  • Number of pages

    9

  • Pages from-to

  • UT code for WoS article

    000332485600013

  • EID of the result in the Scopus database