USB 3.0 readout and time-walk correction method for Timepix3 detector
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216208%3A11110%2F16%3A10337874" target="_blank" >RIV/00216208:11110/16:10337874 - isvavai.cz</a>
Result on the web
<a href="http://dx.doi.org/10.1088/1748-0221/11/12/C12065" target="_blank" >http://dx.doi.org/10.1088/1748-0221/11/12/C12065</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1088/1748-0221/11/12/C12065" target="_blank" >10.1088/1748-0221/11/12/C12065</a>
Alternative languages
Result language
angličtina
Original language name
USB 3.0 readout and time-walk correction method for Timepix3 detector
Original language description
The hybrid particle counting pixel detectors of Medipix family are well known. In this contribution we present new USB 3.0 based interface AdvaDAQ for Timepix3 detector. The AdvaDAQ interface is designed with a maximal emphasis to the flexibility. It is successor of FitPIX interface developed in IEAP CTU in Prague. Its modular architecture supports all Medipix/Timepix chips and all their different readout modes: Medipix2, Timepix (serial and parallel), Medipix3 and Timepix3. The high bandwidth of USB 3.0 permits readout of 1700 full frames per second with Timepix or 8 channel data acquisition from Timepix3 at frequency of 320 MHz. The control and data acquisition is integrated in a multiplatform PiXet software (MS Windows, Mac OS, Linux). In the second part of the publication a new method for correction of the time-walk effect in Timepix3 is described. Moreover, a fully spectroscopic X-ray imaging with Timepix3 detector operated in the ToT mode ( Time-over-Threshold) is presented. It is shown that the AdvaDAQ's readout speed is sufficient to perform spectroscopic measurement at full intensity of radiographic setups equipped with nano- or micro-focus X-ray tubes.
Czech name
—
Czech description
—
Classification
Type
J<sub>x</sub> - Unclassified - Peer-reviewed scientific article (Jimp, Jsc and Jost)
CEP classification
FS - Medical facilities, apparatus and equipment
OECD FORD branch
—
Result continuities
Project
Result was created during the realization of more than one project. More information in the Projects tab.
Continuities
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Others
Publication year
2016
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Name of the periodical
Journal of Instrumentation
ISSN
1748-0221
e-ISSN
—
Volume of the periodical
11
Issue of the periodical within the volume
December
Country of publishing house
GB - UNITED KINGDOM
Number of pages
10
Pages from-to
—
UT code for WoS article
000395731900065
EID of the result in the Scopus database
2-s2.0-85008155536