Evaluation of later morphologic alterations in renal artery wall and renal nerves in response to catheter-based renal denervation in sheep: comparison of the single-point and multiple-point ablation catheters
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216208%3A11140%2F18%3A10384444" target="_blank" >RIV/00216208:11140/18:10384444 - isvavai.cz</a>
Alternative codes found
RIV/00098892:_____/18:N0000122 RIV/00023001:_____/18:00077442
Result on the web
<a href="http://www.biomed.cas.cz/physiolres/pdf/67/67_891.pdf" target="_blank" >http://www.biomed.cas.cz/physiolres/pdf/67/67_891.pdf</a>
DOI - Digital Object Identifier
—
Alternative languages
Result language
angličtina
Original language name
Evaluation of later morphologic alterations in renal artery wall and renal nerves in response to catheter-based renal denervation in sheep: comparison of the single-point and multiple-point ablation catheters
Original language description
This study evaluated the subacute morphologic alterations in renal artery wall and renal nerves in response to catheter-based renal denervation (RDN) in sheep and also compared the efficiency of single-point and multiple-point ablation catheters. Effect of each ablation catheter approved for the clinical use (Symplicity Flex(TM), Medtronic, Inc., or EnligHTN(TM), St. Jude Medical, INC.) was compared to intact contralateral renal artery in 12 sheep by histopathology and immunohistochemistry evaluation after a 10-day period post-RDN procedure. The safety was verified by extensive evaluation of kidney morphology. Vascular wall lesions and nerve injuries were more pronounced in those animals treated with multi-point EnligHTN catheter when compared with animals treated with single-point Symplicity Flex catheter. However, neither RDN procedure led to complete renal nerve ablation. Both systems, tested in the present study, provided only incomplete renal nerve ablation in sheep. Moreover, no appreciable progression of the nerve disintegration in subacute phase post-RDN procedure was observed. This study further supports the notion that the effectiveness remains fully dependent on anatomical inter-individual variability of the sympathetic nerve plexus accompanying the renal artery. Therefore, new systems providing deeper penetrance to targeted perivascular structure would be more efficient.
Czech name
—
Czech description
—
Classification
Type
J<sub>imp</sub> - Article in a specialist periodical, which is included in the Web of Science database
CEP classification
—
OECD FORD branch
30201 - Cardiac and Cardiovascular systems
Result continuities
Project
<a href="/en/project/NV15-34123A" target="_blank" >NV15-34123A: Comparison of the safety and effectiveness of renal denervation performance in sheep using different technical systems.</a><br>
Continuities
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Others
Publication year
2018
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Name of the periodical
Physiological Research
ISSN
0862-8408
e-ISSN
—
Volume of the periodical
67
Issue of the periodical within the volume
6
Country of publishing house
CZ - CZECH REPUBLIC
Number of pages
11
Pages from-to
891-901
UT code for WoS article
000453554300007
EID of the result in the Scopus database
2-s2.0-85058899398