Lab-in-syringe automated protein precipitation and salting-out homogenous liquid-liquid extraction coupled online to UHPLC-MS/MS for the determination of beta-blockers in serum
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216208%3A11160%2F23%3A10471781" target="_blank" >RIV/00216208:11160/23:10471781 - isvavai.cz</a>
Result on the web
<a href="https://verso.is.cuni.cz/pub/verso.fpl?fname=obd_publikace_handle&handle=Y43U.a7P4T" target="_blank" >https://verso.is.cuni.cz/pub/verso.fpl?fname=obd_publikace_handle&handle=Y43U.a7P4T</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1016/j.aca.2023.340966" target="_blank" >10.1016/j.aca.2023.340966</a>
Alternative languages
Result language
angličtina
Original language name
Lab-in-syringe automated protein precipitation and salting-out homogenous liquid-liquid extraction coupled online to UHPLC-MS/MS for the determination of beta-blockers in serum
Original language description
A sample preparation method involving tandem implementation of protein precipitation and salting-out homogenous liquid-liquid extraction was developed for the determination of beta-blockers in serum. The entire procedure was automated using a computer-controlled syringe pump following the Lab-In-Syringe approach. It is based on the denaturation of serum proteins with acetonitrile followed by salt-induced phase separation upon which the proteins accumulate as a compact layer at the interphase of the solutions. The extract is then separated and diluted in-syringe before being submitted to online coupled UHPLC-MS/MS. A 1 mL glass syringe containing a small stir bar for solution mixing at up to 3000 rpm, was used to deal with sample volumes as small as 100 mu L. A sample throughput of 7 h-1 was achieved by performing the chromatographic run and sample preparation procedure in parallel. Linear working ranges were obtained for all analytes between 5 and 100 ng mL-1, with LOD values ranging from 0.4 to 1.5 ng mL-1. Accuracy values in the range of 88.2-106% and high precision of <11% RSD suggest applicability for routine analysis that can be further improved using deuterated standards.
Czech name
—
Czech description
—
Classification
Type
J<sub>imp</sub> - Article in a specialist periodical, which is included in the Web of Science database
CEP classification
—
OECD FORD branch
30104 - Pharmacology and pharmacy
Result continuities
Project
<a href="/en/project/EF15_003%2F0000465" target="_blank" >EF15_003/0000465: Establishment of Specialized Team for Advanced Research on Separation Science</a><br>
Continuities
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)<br>I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Others
Publication year
2023
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Name of the periodical
Analytica Chimica Acta
ISSN
0003-2670
e-ISSN
1873-4324
Volume of the periodical
1251
Issue of the periodical within the volume
April
Country of publishing house
NL - THE KINGDOM OF THE NETHERLANDS
Number of pages
10
Pages from-to
340966
UT code for WoS article
000995347900001