Determination of Indoxyl Sulphate on Carbon Film Composite Electrode and Carbon Paste Electrode
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216208%3A11310%2F18%3A10383431" target="_blank" >RIV/00216208:11310/18:10383431 - isvavai.cz</a>
Result on the web
<a href="https://doi.org/10.2174/1573411013666170718095203" target="_blank" >https://doi.org/10.2174/1573411013666170718095203</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.2174/1573411013666170718095203" target="_blank" >10.2174/1573411013666170718095203</a>
Alternative languages
Result language
angličtina
Original language name
Determination of Indoxyl Sulphate on Carbon Film Composite Electrode and Carbon Paste Electrode
Original language description
Background: This paper is focused on application of a recently developed disposable array of carbon composite electrodes and carbon paste electrode for the determination of indoxyl sulphate. Results: Measurement of basic electrochemical behaviour of indoxyl sulphate on working electrode resulted in the selection of BR buffer pH 2.0 and pH 3.0 as the optimum medium for carbon composite film electrode and carbon paste electrode, respectively. In this medium, concentration dependences of indoxyl sulphate were measured; calculated quantification limits were 0.72 mu mol L-1 for the carbon composite film electrode and 1.7 mu mol L-1 for the carbon paste electrode. Pasivation of the electrodes during repeated measurements and negligible indoxyl sulphate accumulation was observed. Applicability of the method for the indoxyl sulphate determination in spiked urine samples was tested in combination with the solid phase extraction, where methanol, BR buffer, deionised water and their combination were used as elution reagents. Conclusion: Selected graphite-based working electrodes were successfully applied for the development of suitable voltammetric method for the determination of indoxyl sulphate in BR buffer. Both electrodes showed similar behaviour and sufficient sensitivity and accuracy was reached; however, solid phase extraction was not successful due to ineffectual removal of interferents from human urine.
Czech name
—
Czech description
—
Classification
Type
J<sub>imp</sub> - Article in a specialist periodical, which is included in the Web of Science database
CEP classification
—
OECD FORD branch
10406 - Analytical chemistry
Result continuities
Project
<a href="/en/project/GBP206%2F12%2FG151" target="_blank" >GBP206/12/G151: Center of novel approaches to bioanalysis and molecular diagnostics</a><br>
Continuities
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Others
Publication year
2018
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Name of the periodical
Current Analytical Chemistry
ISSN
1573-4110
e-ISSN
—
Volume of the periodical
14
Issue of the periodical within the volume
5
Country of publishing house
NL - THE KINGDOM OF THE NETHERLANDS
Number of pages
6
Pages from-to
530-535
UT code for WoS article
000446494800013
EID of the result in the Scopus database
2-s2.0-85053240617