A Comparative Analysis of Socon CS and Socon Pedicle Screws in View of their Use for Treatment of Osteoporotic Fractures of theThoracolumbal Spine. A Biomechanical Study
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216224%3A14110%2F11%3A00053978" target="_blank" >RIV/00216224:14110/11:00053978 - isvavai.cz</a>
Alternative codes found
RIV/00216305:26210/11:PU93758
Result on the web
—
DOI - Digital Object Identifier
—
Alternative languages
Result language
čeština
Original language name
Srovnávací analýza šroubů Socon CS a Socon v ošetření osteoporotických zlomenin torakolumbální páteře - biomechanická studie
Original language description
PURPOSE OF THE STUDY To compare the anchorage strength of Socon CS cannulated pedicle screws (B. Braun, Aesculap) with that of Socon screws in human cadaver vertebrae, using pull-out strength testing. Twelve samples involving T12 to L3 vertebrae were prepared and a total of 20 pedicle screws, 10 Socon CS and 10 Socon screws, were inserted in them. All screws were 55 mm in length and 6 mm in diameter. After placement, Socon CS pedicle screws were augmented with bone cement (Cimplant cement application kit, B. Braun, Aesculap). The poor quality of osteoporotic bone is responsible for a higher frequency of implant failure due to loosening, particularly when implants developed for healthy bone are used. In this biomechanical study, we tested one of the possibilities of how to reduce the risk of implant failure by pedicle screw augmentation with bone cement.
Czech name
Srovnávací analýza šroubů Socon CS a Socon v ošetření osteoporotických zlomenin torakolumbální páteře - biomechanická studie
Czech description
PURPOSE OF THE STUDY To compare the anchorage strength of Socon CS cannulated pedicle screws (B. Braun, Aesculap) with that of Socon screws in human cadaver vertebrae, using pull-out strength testing. Twelve samples involving T12 to L3 vertebrae were prepared and a total of 20 pedicle screws, 10 Socon CS and 10 Socon screws, were inserted in them. All screws were 55 mm in length and 6 mm in diameter. After placement, Socon CS pedicle screws were augmented with bone cement (Cimplant cement application kit, B. Braun, Aesculap). The poor quality of osteoporotic bone is responsible for a higher frequency of implant failure due to loosening, particularly when implants developed for healthy bone are used. In this biomechanical study, we tested one of the possibilities of how to reduce the risk of implant failure by pedicle screw augmentation with bone cement.
Classification
Type
J<sub>x</sub> - Unclassified - Peer-reviewed scientific article (Jimp, Jsc and Jost)
CEP classification
FJ - Surgery including transplantology
OECD FORD branch
—
Result continuities
Project
—
Continuities
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Others
Publication year
2011
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Name of the periodical
Acta Chirurgiae Orthopaedicae et Traumatologiae Čechoslovaca
ISSN
0001-5415
e-ISSN
—
Volume of the periodical
78
Issue of the periodical within the volume
4
Country of publishing house
CZ - CZECH REPUBLIC
Number of pages
5
Pages from-to
334-338
UT code for WoS article
000295191900008
EID of the result in the Scopus database
—