Reliability of the Ceramic Head of the Total Hip Joint Endoprosthesis Using Weibull's Weakest-link Theory
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216305%3A26210%2F06%3APU63159" target="_blank" >RIV/00216305:26210/06:PU63159 - isvavai.cz</a>
Alternative codes found
RIV/61388998:_____/06:00044148
Result on the web
—
DOI - Digital Object Identifier
—
Alternative languages
Result language
angličtina
Original language name
Reliability of the Ceramic Head of the Total Hip Joint Endoprosthesis Using Weibull's Weakest-link Theory
Original language description
The paper deals with calculation of the parameters of ceramic material from a set of destruction tests of ceramic heads of total hip joint endoprosthesis. The standard way of calculation of the material parameters consists in carrying out a set of 3 or 4point bending tests of specimens cut out from parts of the ceramic material to be analysed. In case of ceramic heads, it is not possible to cut out specimens of required dimensions because the heads are too small (if the cut out specimens were smaller than the normalised ones, the material parameters derived from them would exhibit higher strength values than those which the given ceramic material really has). On that score, a special testing jig was made, in which 40 heads were destructed. From the measured values of circumferential strains of the head's external spherical surface under destruction, the state of stress in the head under destruction was established using the final elements method (FEM). From the values obtained, the so
Czech name
Spolehlivost keramické hlavice totální kyčelní endoporotézy určená pomocí Weibullovy teorie nejslabšího článku
Czech description
Příspěvek se zabývá určením materiálových parametrů keramiky pomocí destrukčních zkoušek keramických hlavic totální kyčelní endoprotézy a jejího následného stochastického zpracování.
Classification
Type
D - Article in proceedings
CEP classification
BO - Biophysics
OECD FORD branch
—
Result continuities
Project
<a href="/en/project/GP101%2F04%2FP037" target="_blank" >GP101/04/P037: Enhancement of the computational modelling level for evaluation of the failure probability of the hip joint endoprosthesis ceramic head</a><br>
Continuities
S - Specificky vyzkum na vysokych skolach
Others
Publication year
2006
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Article name in the collection
World Congress on Medical Physics and Biomedical engineering 2006
ISBN
3-540-36839-6
ISSN
—
e-ISSN
—
Number of pages
4
Pages from-to
2818-2821
Publisher name
Springer
Place of publication
Seoul
Event location
Seoul
Event date
Aug 27, 2006
Type of event by nationality
WRD - Celosvětová akce
UT code for WoS article
—