Design and Fabrication of Planar Magnetoinductive Resonator Arrays for MRI System Field Shaping
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216305%3A26220%2F11%3APU92007" target="_blank" >RIV/00216305:26220/11:PU92007 - isvavai.cz</a>
Result on the web
—
DOI - Digital Object Identifier
—
Alternative languages
Result language
angličtina
Original language name
Design and Fabrication of Planar Magnetoinductive Resonator Arrays for MRI System Field Shaping
Original language description
For the experiments with MR system sensitivity increasing by means of periodical metamaterial structures a magnetoinductive resonator array has to be designed. Proposed design guidelines for precise new array assembly are described and the results of newarrays measurement are presented. For the array's resonant frequency measurement the dip meter method is described. For resonant frequency measurement of single loop a return loss method is presented. As a suitable alternative a new coaxial shunt measurement method is proposed. The results of the new arrays measurement by means of coaxial shunt method are presented also.
Czech name
—
Czech description
—
Classification
Type
J<sub>x</sub> - Unclassified - Peer-reviewed scientific article (Jimp, Jsc and Jost)
CEP classification
JA - Electronics and optoelectronics
OECD FORD branch
—
Result continuities
Project
—
Continuities
S - Specificky vyzkum na vysokych skolach
Others
Publication year
2011
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Name of the periodical
Progress In Electromagnetics
ISSN
1559-9450
e-ISSN
—
Volume of the periodical
2011
Issue of the periodical within the volume
2011
Country of publishing house
US - UNITED STATES
Number of pages
5
Pages from-to
643-647
UT code for WoS article
—
EID of the result in the Scopus database
—