Sound Pressure Level Meter for Ultrasonic Actuators

The result's identifiers

  • Result code in IS VaVaI

    <a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216305%3A26620%2F18%3APR33296" target="_blank" >RIV/00216305:26620/18:PR33296 - isvavai.cz</a>

  • Result on the web

    <a href="http://silense.ceitec.cz/vysledky/UAT" target="_blank" >http://silense.ceitec.cz/vysledky/UAT</a>

  • DOI - Digital Object Identifier

Alternative languages

  • Result language

    angličtina

  • Original language name

    Sound Pressure Level Meter for Ultrasonic Actuators

  • Original language description

    A sound pressure level meter for ultrasonic actuators (SPLM) based on NI's cRIO platform was developed. A rugged controller cRIO-9063 was used both to generate the input signal to the actuator and to sample the measured response. The SPLM implements a master-slave design pattern. A command, generated by an operator is sent to the RT system. RT then prepares a signal of a finite length of desired parameters and triggers the FPGA application, which passes the prepared signal to a Device Under Test (DUT) via its driver. The response of the ultrasonic transducer is measured by a G.R.A.S. pressure microphone and, if available, also the microphone of the DUT. The measured signal is first amplified and then sampled by the digitizer module. Results can be stored in TDMS file format. Additional application was developed to post-process the measured data. Multiple signal types are supported, such as sine, half-sine, square or sweep. Along the sweep signal, a sine signal with a defined frequency step can be generated as well. Amplitude, frequency, frequency change and duration of the generared signal are all user-defined. Additional part of the SPLM is the post-processing application, which can be used to review the measured data. This application calculates the spectral analysis of the measured data, or the sensitivity characteristics of the ultrasonic transducer (if connected). It can also display the sound pressure level response (SPL) of the transducer in tested frequency range.

  • Czech name

  • Czech description

Classification

  • Type

    G<sub>funk</sub> - Functional sample

  • CEP classification

  • OECD FORD branch

    20201 - Electrical and electronic engineering

Result continuities

  • Project

    Result was created during the realization of more than one project. More information in the Projects tab.

  • Continuities

    P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)

Others

  • Publication year

    2018

  • Confidentiality

    S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů

Data specific for result type

  • Internal product ID

    UAT v1.5

  • Numerical identification

    166110

  • Technical parameters

    - rozsah vstupního napětí: ±10 V max - rozsah výstupního napětí: ±10 V max - vzorkovací frekvence (výstup): 1 MS/s - vzorkovací frekvence (vstup): 20 MS/s - počet vstupních kanálů: až 2 - podporované tvary signálu: sinus, half-sinus, obdélník, rozmítaný sinus - tlakový referenční mikrofon: G.R.A.S. 46DE CCP 1/8’’ - frekvenční rozsah (±3 dB): 6.5 – 140 000 Hz - citlivost: 0.79 mV/Pa

  • Economical parameters

    Funčkní vzorek zařízení pro testování ultrazvukových aktuátorů je prozatím používán výhradně pro další výzkum a vývoj, komerční využití se zatím nepředpokládá.

  • Application category by cost

  • Owner IČO

    00216305

  • Owner name

    Vysoké učení technické v Brně

  • Owner country

    CZ - CZECH REPUBLIC

  • Usage type

    V - Výsledek je využíván vlastníkem

  • Licence fee requirement

    A - Poskytovatel licence na výsledek požaduje licenční poplatek

  • Web page

    http://silense.ceitec.cz/vysledky/UAT v1.5