Biomechanical comparison of screw position in variable angle locking plate in intra-articular calcaneal fractures: Cadaveric and radiologic study.

The result's identifiers

  • Result code in IS VaVaI

    <a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F27283933%3A_____%2F25%3AN0000033" target="_blank" >RIV/27283933:_____/25:N0000033 - isvavai.cz</a>

  • Alternative codes found

    RIV/46747885:24410/25:00014560 RIV/46747885:24510/25:00014560 RIV/00216208:11130/25:10498054

  • Result on the web

    <a href="https://www.clinbiomech.com/article/S0268-0033(25)00107-X/abstract" target="_blank" >https://www.clinbiomech.com/article/S0268-0033(25)00107-X/abstract</a>

  • DOI - Digital Object Identifier

    <a href="http://dx.doi.org/10.1016/j.clinbiomech.2025.106534." target="_blank" >10.1016/j.clinbiomech.2025.106534.</a>

Alternative languages

  • Result language

    angličtina

  • Original language name

    Biomechanical comparison of screw position in variable angle locking plate in intra-articular calcaneal fractures: Cadaveric and radiologic study.

  • Original language description

    BACKGROUND: Lateral plating of calcaneal fractures using variable-angle locking plates is still the golden standard for severely comminuted cases. The aim of this study is to explore the possibilities of improving stability of osteosynthesis by changing screw directions. It provides an assessment and comparison of cadaveric biomechanical experiment with retrospective radiologic data analysis. METHODS: In the cadaveric study 8 intact calcaneus-talus specimens were obtained from 4 deceased donors. Fracture type 2b according to Sanders' classification was created in each specimen and fixed with variable-angle locking plate. The specimens were divided in 2 groups differing in orientation of anterior screws and fixed in PMMA base. A push-in test was performed by a two-column testing machine until gross failure. Retrospective cohort study was performed, reviewing data of 74 patients which underwent surgical treatment of calcaneal fractures with the same construct. Evaluation was performed at scheduled CT and X-Ray controls. Direction of inserted screws and implant failure were noted. FINDINGS: The cadaveric study proved that there is no significant difference in mean failure force between two abovementioned screw configurations in Sanders 2b fracture. A significant difference was observed in initial stiffness. The radiologic retrospective study showed that difference in screw position within all fracture types but type 2b is significant. INTERPRETATION: Screw configuration in the anterior part of variable-angle locking plate appears to affect primary stiffness and stability of the construct. Particularly in more comminuted fractures, screw inserted in the direction of sustentaculum improves the stability and lowers risk of implant failure.

  • Czech name

  • Czech description

Classification

  • Type

    J<sub>ost</sub> - Miscellaneous article in a specialist periodical

  • CEP classification

  • OECD FORD branch

    30401 - Health-related biotechnology

Result continuities

  • Project

  • Continuities

    N - Vyzkumna aktivita podporovana z neverejnych zdroju

Others

  • Publication year

    2025

  • Confidentiality

    S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů

Data specific for result type

  • Name of the periodical

    Clin Biomech

  • ISSN

    0268-0033

  • e-ISSN

    1879-1271

  • Volume of the periodical

    125

  • Issue of the periodical within the volume

    23.4.2025

  • Country of publishing house

    GB - UNITED KINGDOM

  • Number of pages

    8

  • Pages from-to

    106534

  • UT code for WoS article

    001483115900001

  • EID of the result in the Scopus database

    2-s2.0-105003132466