CPR Induced Inappropriate Shocks from a Subcutaneous Implantable Cardioverter Defibrillator during Out-of-Hospital Cardiac Arrest
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F44555601%3A13450%2F19%3A43894899" target="_blank" >RIV/44555601:13450/19:43894899 - isvavai.cz</a>
Alternative codes found
RIV/00216208:11120/20:43917896
Result on the web
<a href="https://www.tandfonline.com/doi/full/10.1080/10903127.2019.1599475" target="_blank" >https://www.tandfonline.com/doi/full/10.1080/10903127.2019.1599475</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1080/10903127.2019.1599475" target="_blank" >10.1080/10903127.2019.1599475</a>
Alternative languages
Result language
angličtina
Original language name
CPR Induced Inappropriate Shocks from a Subcutaneous Implantable Cardioverter Defibrillator during Out-of-Hospital Cardiac Arrest
Original language description
We present the case of a 30-year-old man with a subcutaneous implantable cardioverter defibrillator (S-ICD) who suffered sudden out-of-hospital cardiac arrest. During resuscitation, the patient received inappropriate shocks due to oversensing by the S-ICD of chest compression induced artifact. The rhythm on the EMS monitor showed asystole throughout the period of resuscitation. Placement of a magnet over the S-ICD generator failed to inhibit the delivery of S-ICD shocks. Information regarding inappropriate S-ICD shocks may be useful during resuscitation of patients with sudden cardiac arrest.
Czech name
—
Czech description
—
Classification
Type
J<sub>imp</sub> - Article in a specialist periodical, which is included in the Web of Science database
CEP classification
—
OECD FORD branch
30201 - Cardiac and Cardiovascular systems
Result continuities
Project
—
Continuities
N - Vyzkumna aktivita podporovana z neverejnych zdroju
Others
Publication year
2019
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Name of the periodical
Prehospital Emergency Care
ISSN
1090-3127
e-ISSN
—
Volume of the periodical
24
Issue of the periodical within the volume
1
Country of publishing house
US - UNITED STATES
Number of pages
5
Pages from-to
85-89
UT code for WoS article
000466029500001
EID of the result in the Scopus database