Flow electrochemical biosensors based on enzymatic porous reactor and tubular detector of silver solid amalgam
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F61388955%3A_____%2F13%3A00392185" target="_blank" >RIV/61388955:_____/13:00392185 - isvavai.cz</a>
Result on the web
<a href="http://dx.doi.org/10.1016/j.aca.2013.03.055" target="_blank" >http://dx.doi.org/10.1016/j.aca.2013.03.055</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1016/j.aca.2013.03.055" target="_blank" >10.1016/j.aca.2013.03.055</a>
Alternative languages
Result language
angličtina
Original language name
Flow electrochemical biosensors based on enzymatic porous reactor and tubular detector of silver solid amalgam
Original language description
A flow amperometric enzymatic biosensor for the determination of glucose was constructed. The biosensor consists of a flow reactor based on porous silver solid amalgam (AgSA) and a flow tubular detector based on compact AgSA. The preparation of the sensor and the determination of glucose occurred in three steps. First, a self-assembled monolayer of 11-mercaptoundecanoic acid (MUA) was formed at the porous surface of the reactor. Second, enzyme glucose oxidase (GOx) was covalently immobilized at MUA-layer using N-ethyl-N-(3-dimethylaminopropyl) carboimide and N-hydroxysuccinimide chemistry. Finally, a decrease of oxygen concentration (directly proportional to the concentration of glucose) during enzymatic reaction was amperometrically measured on the tubular detector under flow injection conditions. The following parameters of glucose determination were optimized with respect to amperometric response: composition of the mobile phase, its concentration, the potential of detection and the
Czech name
—
Czech description
—
Classification
Type
J<sub>x</sub> - Unclassified - Peer-reviewed scientific article (Jimp, Jsc and Jost)
CEP classification
CF - Physical chemistry and theoretical chemistry
OECD FORD branch
—
Result continuities
Project
<a href="/en/project/GAP206%2F11%2F1638" target="_blank" >GAP206/11/1638: Novel electrochemical sensors and sensing techniques for the analysis of nucleic acids structure and interactions</a><br>
Continuities
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Others
Publication year
2013
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Name of the periodical
Analytica Chimica Acta
ISSN
0003-2670
e-ISSN
—
Volume of the periodical
778
Issue of the periodical within the volume
MAY 2013
Country of publishing house
NL - THE KINGDOM OF THE NETHERLANDS
Number of pages
7
Pages from-to
24-30
UT code for WoS article
000318959400004