A miniaturized electrode system for voltammetric determination of electrochemically reducible environmental pollutants
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F61388955%3A_____%2F16%3A00453603" target="_blank" >RIV/61388955:_____/16:00453603 - isvavai.cz</a>
Alternative codes found
RIV/00216208:11310/16:10321857
Result on the web
<a href="http://dx.doi.org/10.1016/j.snb.2015.11.136" target="_blank" >http://dx.doi.org/10.1016/j.snb.2015.11.136</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1016/j.snb.2015.11.136" target="_blank" >10.1016/j.snb.2015.11.136</a>
Alternative languages
Result language
angličtina
Original language name
A miniaturized electrode system for voltammetric determination of electrochemically reducible environmental pollutants
Original language description
A newly designed miniaturized electrode system (MES) based on a mercury meniscus modified silver solid amalgam electrode as the working electrode was developed for voltammetric determination of electrochemically reducible organic compounds in small sample volumes. Genotoxic 2-aminofluoren-9-one (2-AFN) was chosen as a model environmental pollutant in this study to test the performance of the introduced MES. The determination of 2-AFN was carried out by differential pulse voltammetry in 100 μL samples under following conditions, using two different ways of air oxygen removal: (i) a supporting electrolyte methanol–Britton–Robinson (BR) buffer of pH 4.0 (1:9, v/v), with regeneration potentials Ereg,1 = 0 mV and Ereg,2 = 1200 mV and with oxygen removal by bubbling with nitrogen, and (ii) a supporting electrolyte methanol–BR buffer of pH 10.0 (1:9, v/v), with regeneration potentials Ereg,1 = 200 mV and Ereg,2 = 1600 mV and with oxygen removal by the addition of solid sodium sulfite as a reducing agent. The calibration curves were measured in the concentration range from 1 to 100 μmol L1 of 2-AFN, with the limits of quantification about 1 μmol L1 reached in both media. The practical applicability of the newly developed voltammetric methods using the MES was successfully verified on the direct determination of 2-AFN in model samples of drinking and river water.n
Czech name
—
Czech description
—
Classification
Type
J<sub>x</sub> - Unclassified - Peer-reviewed scientific article (Jimp, Jsc and Jost)
CEP classification
CF - Physical chemistry and theoretical chemistry
OECD FORD branch
—
Result continuities
Project
Result was created during the realization of more than one project. More information in the Projects tab.
Continuities
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Others
Publication year
2016
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Name of the periodical
Sensors and Actuators B - Chemical
ISSN
0925-4005
e-ISSN
—
Volume of the periodical
227
Issue of the periodical within the volume
MAY 2016
Country of publishing house
CH - SWITZERLAND
Number of pages
8
Pages from-to
263-270
UT code for WoS article
000369944200034
EID of the result in the Scopus database
2-s2.0-84952776389