Proposal of electrode for measuring glucose concentration in blood
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F61989100%3A27240%2F19%3A10241646" target="_blank" >RIV/61989100:27240/19:10241646 - isvavai.cz</a>
Result on the web
<a href="https://link.springer.com/chapter/10.1007/978-981-10-9023-3_145" target="_blank" >https://link.springer.com/chapter/10.1007/978-981-10-9023-3_145</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1007/978-981-10-9023-3_145" target="_blank" >10.1007/978-981-10-9023-3_145</a>
Alternative languages
Result language
angličtina
Original language name
Proposal of electrode for measuring glucose concentration in blood
Original language description
The article deals with the design of the electrode system and the ap-propriate measuring circuit for the measurement of blood glucose concentration, since the determination of the level of blood glucose is an integral part of many medical procedures for determining the state of the human organism. The aim of the concept is to create two functional devices with different design of the elec-trode system and to perform a series of measurements to verify the functionality and the data statistically process. There was used the principle of resistive sensing of non-electric quantities, so the designed systems using gold electrodes working as electrolytic sensors. The problem relating to the measuring circuit, which is the same for both electrode systems, has been solved by a microampere meter which can convert the generated signal to a suitable measured electrical signal. The signal is formed by the reaction of glucose and enzyme, the resulting value being directly proportional to the glucose concentration. At this work, created systems fully meet specified conditions and requirements. The research con-firmed the functionality of the designed units. On the basis of obtained data is possible to compare and evaluate the constructed electrode systems and to make the basis for further development in the subject matter.
Czech name
—
Czech description
—
Classification
Type
D - Article in proceedings
CEP classification
—
OECD FORD branch
20201 - Electrical and electronic engineering
Result continuities
Project
<a href="/en/project/GA17-03037S" target="_blank" >GA17-03037S: Investment evaluation of medical device development</a><br>
Continuities
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Others
Publication year
2019
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Article name in the collection
IFMBE Proceedings. Volume 68, issue 3
ISBN
978-981-10-9022-6
ISSN
1680-0737
e-ISSN
—
Number of pages
3
Pages from-to
801-804
Publisher name
Springer
Place of publication
New York
Event location
Praha
Event date
Jun 3, 2018
Type of event by nationality
WRD - Celosvětová akce
UT code for WoS article
000449744300145