A New Approach for Testing Fetal Heart Rate Monitors
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F61989100%3A27240%2F20%3A10245279" target="_blank" >RIV/61989100:27240/20:10245279 - isvavai.cz</a>
Result on the web
<a href="https://www.mdpi.com/1424-8220/20/15/4139" target="_blank" >https://www.mdpi.com/1424-8220/20/15/4139</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.3390/s20154139" target="_blank" >10.3390/s20154139</a>
Alternative languages
Result language
angličtina
Original language name
A New Approach for Testing Fetal Heart Rate Monitors
Original language description
In this paper, a new approach for the periodical testing and the functionality evaluation of a fetal heart rate monitor device based on ultrasound principle is proposed. The design and realization of the device are presented, together with the description of its features and functioning tests. In the designed device, a relay element, driven by an electric signal that allows switching at two specific frequencies, is used to simulate the fetus and the mother's heartbeat. The simulator was designed to be compliant with the standard requirements for accurate assessment and measurement of medical devices. The accuracy of the simulated signals was evaluated, and it resulted to be stable and reliable. The generated frequencies show an error of about 0.5% with respect to the nominal one while the accuracy of the test equipment was within +/- 3% of the test signal set frequency. This value complies with the technical standard for the accuracy of fetal heart rate monitor devices. Moreover, the performed tests and measurements show the correct functionality of the developed simulator. The proposed equipment and testing respect the technical requirements for medical devices. The features of the proposed device make it simple and quick in testing a fetal heart rate monitor, thus providing an efficient way to evaluate and test the correlation capabilities of commercial apparatuses.
Czech name
—
Czech description
—
Classification
Type
J<sub>imp</sub> - Article in a specialist periodical, which is included in the Web of Science database
CEP classification
—
OECD FORD branch
20601 - Medical engineering
Result continuities
Project
<a href="/en/project/TL01000302" target="_blank" >TL01000302: Medical devices development as an effective investment for public and private entities</a><br>
Continuities
S - Specificky vyzkum na vysokych skolach
Others
Publication year
2020
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Name of the periodical
Sensors
ISSN
1424-3210
e-ISSN
—
Volume of the periodical
20
Issue of the periodical within the volume
15
Country of publishing house
CH - SWITZERLAND
Number of pages
15
Pages from-to
—
UT code for WoS article
000559132500001
EID of the result in the Scopus database
—