A Novel FBG-Based Triggering System for Cardiac MR Imaging at 3 Tesla: A Pilot Pre-Clinical Study
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F61989100%3A27240%2F20%3A10245658" target="_blank" >RIV/61989100:27240/20:10245658 - isvavai.cz</a>
Alternative codes found
RIV/00216224:14740/20:00118190
Result on the web
<a href="https://ieeexplore.ieee.org/document/9210616" target="_blank" >https://ieeexplore.ieee.org/document/9210616</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1109/ACCESS.2020.3028224" target="_blank" >10.1109/ACCESS.2020.3028224</a>
Alternative languages
Result language
angličtina
Original language name
A Novel FBG-Based Triggering System for Cardiac MR Imaging at 3 Tesla: A Pilot Pre-Clinical Study
Original language description
This first-ever study demonstrates the applicability of a fiber Bragg grating (FBG) system for MR cardiac triggering of cardiovascular magnetic resonance at 3 Tesla. The unique patented system senses body movements caused by cardiac activity using a non-invasive ballistocardiography (BCG) sensor. The pilot research compares a novel FBG-based system with clinically used triggering systems based on electrocardiography (ECG) and pulse oximetry (POX). The pilot pre-clinical study was conducted on 8 subjects at a Siemens Prisma 3T MR Scanner. The study compares images from two basic cardiac sequences, TRUE FISP (Free Induction Decay Steady-State Precession) and PSIR (Phase Sensitive Inversion Recovery), using objective methods and subjective evaluation by clinical experts. The study presents original results that confirm the applicability of optical sensors in the field of cardiac triggering having a number of advantages in comparison to conventional solutions, such as no eddy current interference, ease of placement of the sensor on the patient's body, and senor reusability. The proposed FBG-based system achieves comparable results with the most frequently used and most accurate ECG-based and POX-based clinical systems. In terms of subjective evaluation by experts, the FBG system outperformed the POX-based system used in clinical practice.
Czech name
—
Czech description
—
Classification
Type
J<sub>imp</sub> - Article in a specialist periodical, which is included in the Web of Science database
CEP classification
—
OECD FORD branch
20201 - Electrical and electronic engineering
Result continuities
Project
Result was created during the realization of more than one project. More information in the Projects tab.
Continuities
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)<br>S - Specificky vyzkum na vysokych skolach
Others
Publication year
2020
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Name of the periodical
IEEE Access
ISSN
2169-3536
e-ISSN
—
Volume of the periodical
8
Issue of the periodical within the volume
2020
Country of publishing house
US - UNITED STATES
Number of pages
19
Pages from-to
181205-181223
UT code for WoS article
000577884000001