A new poweful system to test ultrasonograph quality
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F61989592%3A15110%2F14%3A33153244" target="_blank" >RIV/61989592:15110/14:33153244 - isvavai.cz</a>
Result on the web
<a href="http://dx.doi.org/10.1109/ULTSYM.2014.0302" target="_blank" >http://dx.doi.org/10.1109/ULTSYM.2014.0302</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1109/ULTSYM.2014.0302" target="_blank" >10.1109/ULTSYM.2014.0302</a>
Alternative languages
Result language
angličtina
Original language name
A new poweful system to test ultrasonograph quality
Original language description
Measuring equipment mapping a Point Spread Function distribution over scanned area to precisely evaluate nine distinctive quality parameters of the whole ultrasound imaging system using one measuring process was designed. The paper describes the method of measurement and specifies the obtained parameters including brief data of their uncertainty study. The method is dedicated mainly for complex tests of standard B-mode imaging systems using any type of transducer (including 3D) in frequency range 1 - 30MHz.
Czech name
—
Czech description
—
Classification
Type
D - Article in proceedings
CEP classification
FS - Medical facilities, apparatus and equipment
OECD FORD branch
—
Result continuities
Project
—
Continuities
O - Projekt operacniho programu
Others
Publication year
2014
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Article name in the collection
Ultrasonics Symposium (IUS), 2014 IEEE International
ISBN
978-1-4799-7049-0
ISSN
—
e-ISSN
—
Number of pages
4
Pages from-to
1225-1228
Publisher name
IEEE
Place of publication
Neuveden
Event location
Chicago, IL
Event date
Sep 3, 2014
Type of event by nationality
WRD - Celosvětová akce