High sensitivity deep-UV LED-based z-cell photometric detector for capillary liquid chromatography
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F62156489%3A43210%2F18%3A43914286" target="_blank" >RIV/62156489:43210/18:43914286 - isvavai.cz</a>
Alternative codes found
RIV/00216305:26620/18:PU131551
Result on the web
<a href="https://doi.org/10.1016/j.aca.2018.06.005" target="_blank" >https://doi.org/10.1016/j.aca.2018.06.005</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1016/j.aca.2018.06.005" target="_blank" >10.1016/j.aca.2018.06.005</a>
Alternative languages
Result language
angličtina
Original language name
High sensitivity deep-UV LED-based z-cell photometric detector for capillary liquid chromatography
Original language description
A new high sensitivity deep-UV LED photometric detector with a z-type flow cell (45 nL or 180 nL) for miniaturised and portable capillary liquid chromatography (LC) was designed and fabricated to overcome sensitivity limitations due to short pathlength in on-capillary detectors. The new detector has a 10 mm geometric pathlength and uses high intensity light-emitting diodes (LED) as light sources in the deep-UV range (254 nm and 280 nm). No optical reference was necessary due to the low drift in the signal. Stray light was minimized by the use of an adjustable slit with a 0.5mm pinhole. The direct relationship between absorbance and concentration was obtained using dichromate to evaluate the sensitivity and the linearity range of the detector. Performance of the miniaturised version was compared with that obtained from a commercial benchtop detector for capillary LC under the same conditions using the same optical z-cell. The miniaturised version exhibited a superior performance across all parameters, including 3 times higher effective pathlength, 4 times higher upper limit of detector linearity, and 2-5 times lower stray light levels. An application of the new detector was shown with the detection of L-dopa, L-tyrosine, norfenefrine, phenylephrine and tyramine, separated using capillary LC. The baseline noise level recorded was as low as 3.9 mu AU. Further, the detector was applied in a miniaturised capillary LC for the indirect detection of common inorganic anions. In comparison to an on-capillary LED detector applied under similar chromatographic conditions, there was a 50 times higher signal to noise (S/N) ratio.
Czech name
—
Czech description
—
Classification
Type
J<sub>imp</sub> - Article in a specialist periodical, which is included in the Web of Science database
CEP classification
—
OECD FORD branch
10406 - Analytical chemistry
Result continuities
Project
—
Continuities
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Others
Publication year
2018
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Name of the periodical
Analytica Chimica Acta
ISSN
0003-2670
e-ISSN
—
Volume of the periodical
1032
Issue of the periodical within the volume
22 November
Country of publishing house
NL - THE KINGDOM OF THE NETHERLANDS
Number of pages
6
Pages from-to
197-202
UT code for WoS article
000443002100022
EID of the result in the Scopus database