Comparison of Benchtop and Portable Near-Infrared Instruments to Predict the Type of Microplastic Added to High-Moisture Food Samples
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F62157124%3A16270%2F25%3A43882861" target="_blank" >RIV/62157124:16270/25:43882861 - isvavai.cz</a>
Result on the web
<a href="https://www.mdpi.com/1424-8220/26/1/210" target="_blank" >https://www.mdpi.com/1424-8220/26/1/210</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.3390/s26010210" target="_blank" >10.3390/s26010210</a>
Alternative languages
Result language
angličtina
Original language name
Comparison of Benchtop and Portable Near-Infrared Instruments to Predict the Type of Microplastic Added to High-Moisture Food Samples
Original language description
Near-infrared (NIR) spectroscopy is a rapid, non-destructive analytical tool widely used in the food and agricultural sectors. In this study, two NIR instruments were compared for classifying the addition of microplastics (MPs) to high-moisture-content samples such as vegetables and fruit. Polyethylene (PE), polypropylene (PP), and a mix of polymers (PE + PP) MP were added to mixtures of spinach and banana and scanned using benchtop (Bruker Tango) and portable (MicroNIR) instruments. Both principal component analysis (PCA) and partial least squares (PLS) were used to analyze and interpret the spectra of the samples. Quantitative models were developed to predict the addition of Mix, PP, or PE to spinach and banana samples using PLS regression. The R2 CV and the SECV obtained were 0.88 and 0.44 for the benchtop samples, and 0.54 and 0.67 for the portable instruments, respectively. Two wavenumber regions were also evaluated: 11,520-7500 cm-1 (short to medium wavelengths), and 7500-4200 cm-1 (long wavelengths). The R2 CV and the SECV obtained were 0.88 and 0.46, 0.86 and 0.49, respectively, for the prediction of addition in samples analyzed on the benchtop instrument using short and long wavenumbers, respectively. This study provides new insights into the comparison of two instruments for detecting the addition of MPs in high-moisture samples. The results of this study will ensure that NIR can be utilized not only to measure the quality of these samples but also to monitor MPs.
Czech name
—
Czech description
—
Classification
Type
J<sub>imp</sub> - Article in a specialist periodical, which is included in the Web of Science database
CEP classification
—
OECD FORD branch
10406 - Analytical chemistry
Result continuities
Project
—
Continuities
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Others
Publication year
2025
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Name of the periodical
Sensors
ISSN
1424-8220
e-ISSN
1424-8220
Volume of the periodical
26
Issue of the periodical within the volume
1
Country of publishing house
CH - SWITZERLAND
Number of pages
11
Pages from-to
nestránkováno
UT code for WoS article
001657620500001
EID of the result in the Scopus database
—