Pressure Drop in Microfluidic Reactor.
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F67985858%3A_____%2F17%3A00473884" target="_blank" >RIV/67985858:_____/17:00473884 - isvavai.cz</a>
Result on the web
<a href="http://hdl.handle.net/11104/0270998" target="_blank" >http://hdl.handle.net/11104/0270998</a>
DOI - Digital Object Identifier
—
Alternative languages
Result language
angličtina
Original language name
Pressure Drop in Microfluidic Reactor.
Original language description
Microfluidic reactors are devices of internal volume in few microliters, which can be exploited to conduct complex reactions in a continuous mode. In this study, the Labtrix START (Chemtrix, NL) platform with a glass microfluidic reactor was examined from the hydrodynamics point of view for planned reaction – a stereoselective hydrogenation. The comparison of the measured and predicted pressure drop in the microfluidic reactor indicated that the measured pressure drop was about two times higher than the theoretical prediction.