Calibration of germanium temperature sensors of helium cryostat for experimental study of natural turbulent convection
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68081731%3A_____%2F10%3A00352300" target="_blank" >RIV/68081731:_____/10:00352300 - isvavai.cz</a>
Result on the web
—
DOI - Digital Object Identifier
—
Alternative languages
Result language
angličtina
Original language name
Calibration of germanium temperature sensors of helium cryostat for experimental study of natural turbulent convection
Original language description
Experimental study of Nusselt number on Rayleigh number dependence with using of cryogenic 4He gas requires precise measurement of the temperature differenceT between the top and the bottom plates of the convection cell. A special calibration device wasadapted and used for comparative calibration of germanium temperature sensors. This device enables to compare up to four sensors within the temperature range of 4.3 K ? 40 K.
Czech name
—
Czech description
—
Classification
Type
D - Article in proceedings
CEP classification
BJ - Thermodynamics
OECD FORD branch
—
Result continuities
Project
Result was created during the realization of more than one project. More information in the Projects tab.
Continuities
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)<br>Z - Vyzkumny zamer (s odkazem do CEZ)
Others
Publication year
2010
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Article name in the collection
The 11th Cryogenics 2010 - IIR International Conference
ISBN
978-2-913149-77-9
ISSN
—
e-ISSN
—
Number of pages
3
Pages from-to
—
Publisher name
ICARIS Ltd
Place of publication
Praha
Event location
Bratislava
Event date
Apr 26, 2010