Water Bolus Influence on Temperature Distribution of Applicator for Small Melanoma Tumors
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21230%2F09%3A00158065" target="_blank" >RIV/68407700:21230/09:00158065 - isvavai.cz</a>
Result on the web
—
DOI - Digital Object Identifier
—
Alternative languages
Result language
angličtina
Original language name
Water Bolus Influence on Temperature Distribution of Applicator for Small Melanoma Tumors
Original language description
In this paper influence of the water bolus temperature on the thermal distribution inside an homogeneous muscles phantom are analyzed. Temperature simulations were used for the analysis considering two different thermal models. The results show dependence of the temperature peak deepness which position reduces increasing the bolus temperature. These results can be used as a preliminary guideline water bolus temperature choise during in-vivo experiments.
Czech name
—
Czech description
—
Classification
Type
D - Article in proceedings
CEP classification
JA - Electronics and optoelectronics
OECD FORD branch
—
Result continuities
Project
<a href="/en/project/GD102%2F08%2FH081" target="_blank" >GD102/08/H081: Non-standard application of physical fields - analogy, modelling, verification and simulation</a><br>
Continuities
Z - Vyzkumny zamer (s odkazem do CEZ)
Others
Publication year
2009
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Article name in the collection
European Microwave Week 2009 - Conference Proceedings
ISBN
978-1-4244-4748-0
ISSN
—
e-ISSN
—
Number of pages
4
Pages from-to
—
Publisher name
Horizon House Publications
Place of publication
London
Event location
Rome
Event date
Sep 28, 2009