Calibration of Rogowski Coils with an Integrator at High Currents

The result's identifiers

  • Result code in IS VaVaI

    <a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21230%2F11%3A00181968" target="_blank" >RIV/68407700:21230/11:00181968 - isvavai.cz</a>

  • Alternative codes found

    RIV/00177016:_____/11:#0000424

  • Result on the web

    <a href="http://dx.doi.org/10.1109/TIM.2011.2126090" target="_blank" >http://dx.doi.org/10.1109/TIM.2011.2126090</a>

  • DOI - Digital Object Identifier

    <a href="http://dx.doi.org/10.1109/TIM.2011.2126090" target="_blank" >10.1109/TIM.2011.2126090</a>

Alternative languages

  • Result language

    angličtina

  • Original language name

    Calibration of Rogowski Coils with an Integrator at High Currents

  • Original language description

    This paper describes the calibration of a Rogowski coil (RC) with an integrator assigned for current measurement in the range up to 30 kA at 50 Hz mains frequency. A current loop with 10 turns with a current of 3 kA was used for calibration. The magneticfield generated in this way differs from the field of one turn placed in the centre and passed with a current of 30 kA. The influence of unevenness of winding and the cross-section of the RC may cause a calibration error when using a current loop. The difference in the results between using a current loop and one turn and the measurement uncertainties are given below. Calculations of the mutual inductances of typical RCs and their uncertainties are presented in the Appendix.

  • Czech name

  • Czech description

Classification

  • Type

    J<sub>x</sub> - Unclassified - Peer-reviewed scientific article (Jimp, Jsc and Jost)

  • CEP classification

    JB - Sensors, detecting elements, measurement and regulation

  • OECD FORD branch

Result continuities

  • Project

  • Continuities

    Z - Vyzkumny zamer (s odkazem do CEZ)

Others

  • Publication year

    2011

  • Confidentiality

    S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů

Data specific for result type

  • Name of the periodical

    IEEE Transactions on Instrumentation and Measurement

  • ISSN

    0018-9456

  • e-ISSN

  • Volume of the periodical

    60

  • Issue of the periodical within the volume

    7

  • Country of publishing house

    US - UNITED STATES

  • Number of pages

    5

  • Pages from-to

    2434-2438

  • UT code for WoS article

    000291432000045

  • EID of the result in the Scopus database