Calibration of the 3D Coil System's Orthogonalities
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21230%2F12%3A00195000" target="_blank" >RIV/68407700:21230/12:00195000 - isvavai.cz</a>
Result on the web
—
DOI - Digital Object Identifier
—
Alternative languages
Result language
angličtina
Original language name
Calibration of the 3D Coil System's Orthogonalities
Original language description
Two methods were tested for calibration procedures of the 3D coil system which is used as a source part of the magnetic tracker. Precise magnetometer's positions with respect to the coil system were used for determining of the non-orthogonality of the coil system as well as for determining of the magnetic moment. The needed number of the reference positions was measured to calculate six unknown calibrated parameters. In comparison, the 3D coils system was also calibrated with a method which employs theorthogonal homogenous AC field generated in Helmholtz coils. The misalignment of up to two degrees was discovered.