Uncertainty Analysis of Calibration of the 3D Coil System
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21230%2F12%3A00199400" target="_blank" >RIV/68407700:21230/12:00199400 - isvavai.cz</a>
Result on the web
<a href="http://iris.elf.stuba.sk/cgi-bin/jeeec?act=pr&no=7s_112" target="_blank" >http://iris.elf.stuba.sk/cgi-bin/jeeec?act=pr&no=7s_112</a>
DOI - Digital Object Identifier
—
Alternative languages
Result language
angličtina
Original language name
Uncertainty Analysis of Calibration of the 3D Coil System
Original language description
The scalar calibration of the 3D coil system can be used as an efficient metrological method to identify all coil system's parame- ters. To express the method accuracy, the uncertainty of sensitivity establishing has been analytically calculated and theuncertainty of the orthogonal angles, which are numerically calculated parameters, , was evaluated with the Monte Carlo method. The method uncertainty was affected mostly by the calibration current stability and current measurements.
Czech name
—
Czech description
—
Classification
Type
J<sub>x</sub> - Unclassified - Peer-reviewed scientific article (Jimp, Jsc and Jost)
CEP classification
JB - Sensors, detecting elements, measurement and regulation
OECD FORD branch
—
Result continuities
Project
Result was created during the realization of more than one project. More information in the Projects tab.
Continuities
S - Specificky vyzkum na vysokych skolach
Others
Publication year
2012
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Name of the periodical
Journal of Electrical Engineering
ISSN
1335-3632
e-ISSN
—
Volume of the periodical
63
Issue of the periodical within the volume
December
Country of publishing house
SK - SLOVAKIA
Number of pages
4
Pages from-to
90-93