Timepix3: Temperature Influence on Radiation Energy Measurement with Si Sensor
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21230%2F23%3A00363682" target="_blank" >RIV/68407700:21230/23:00363682 - isvavai.cz</a>
Result on the web
<a href="https://doi.org/10.3390/s23042201" target="_blank" >https://doi.org/10.3390/s23042201</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.3390/s23042201" target="_blank" >10.3390/s23042201</a>
Alternative languages
Result language
angličtina
Original language name
Timepix3: Temperature Influence on Radiation Energy Measurement with Si Sensor
Original language description
The Timepix3 readout ASIC chip is a hybrid pixelated radiation detector, designed at CERN, which contains a 256 pxx256 px matrix. Each of the 65,536 radiation-sensitive pixels can record an incoming particle, its energy deposition or time of arrival and measure them simultaneously. Since the detector is suitable for a wide range of applications from particle physics, national security and medicine to space science, it can be used in a wide range of temperatures. Until now, it has to be calibrated every time to the operating point of the application. This paper studies the possibility of energy measurement with Timepix3 equipped with a 500 µm thick silicon sensor and MiniPIX readout interface in the temperatures between 10 °C and 70 °C with only one calibration. The detector has been irradiated by X-ray fluorescence photons in the energy range from 8 keV to 57 keV, and 31 keV to 81 keV photons from the 133Ba radioactive source. A deviation of 5% in apparent energy value may occur for a 10 °C change in temperature from the reference point, but, with the next temperature change, it can reach up to -30%. Moreover, Barium photons with an energy of 81 keV appear as deposited energy of only 55 keV at a detector temperature of 70 °C. An original compensation method that reduces the relative measurement error from -30% to less than 1% is presented in this paper.
Czech name
—
Czech description
—
Classification
Type
J<sub>imp</sub> - Article in a specialist periodical, which is included in the Web of Science database
CEP classification
—
OECD FORD branch
10304 - Nuclear physics
Result continuities
Project
<a href="/en/project/EF16_019%2F0000778" target="_blank" >EF16_019/0000778: Center for advanced applied science</a><br>
Continuities
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)<br>S - Specificky vyzkum na vysokych skolach
Others
Publication year
2023
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Name of the periodical
Sensors
ISSN
1424-8220
e-ISSN
1424-8220
Volume of the periodical
23
Issue of the periodical within the volume
4
Country of publishing house
CH - SWITZERLAND
Number of pages
14
Pages from-to
—
UT code for WoS article
000942406500001
EID of the result in the Scopus database
2-s2.0-85149168457