Study of Temperature Influence on Timepix2 Energy Measurements

The result's identifiers

  • Result code in IS VaVaI

    <a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21230%2F25%3A00388247" target="_blank" >RIV/68407700:21230/25:00388247 - isvavai.cz</a>

  • Alternative codes found

    RIV/68407700:21670/25:00388247

  • Result on the web

    <a href="https://doi.org/10.1088/1748-0221/20/11/C11003" target="_blank" >https://doi.org/10.1088/1748-0221/20/11/C11003</a>

  • DOI - Digital Object Identifier

    <a href="http://dx.doi.org/10.1088/1748-0221/20/11/C11003" target="_blank" >10.1088/1748-0221/20/11/C11003</a>

Alternative languages

  • Result language

    angličtina

  • Original language name

    Study of Temperature Influence on Timepix2 Energy Measurements

  • Original language description

    In this study, we examine the temperature dependence of a Timepix2 (v1) detector (256x 256 pixels, 55 μm pixel pitch) coupled to a 500 μm thick silicon sensor. Measurements were performed with γ-photons 59.5 keV and α-particles 5.5 MeV in vacuum over a stabilized temperature range of -20 °C to 80 °C. Two different readout systems were used to verify that the response of the Timepix2 detector was not affected by the readout electronics. In both normal and adaptive gain modes, the detector demonstrated excellent stability in Time-over-Threshold (ToT) based energy measurements, with relative deviations of peak positions remaining below 4% across the temperature range from -20 °C to 80 °C. Digital-to-Analog Converter (DAC) voltage measurements revealed a significant temperature dependence of the DAC Ikrum, which compensates for threshold shifts and enables accurate ToT readings. In adaptive gain mode, a systematic shift of approximately 0.6% per 10 °C was observed in absolute ToT accuracy.

  • Czech name

  • Czech description

Classification

  • Type

    J<sub>imp</sub> - Article in a specialist periodical, which is included in the Web of Science database

  • CEP classification

  • OECD FORD branch

    20201 - Electrical and electronic engineering

Result continuities

  • Project

    <a href="/en/project/GM23-04869M" target="_blank" >GM23-04869M: Particle identification in high-energy physics experiments and space with advanced detection systems</a><br>

  • Continuities

    P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)

Others

  • Publication year

    2025

  • Confidentiality

    S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů

Data specific for result type

  • Name of the periodical

    Journal of Instrumentation

  • ISSN

    1748-0221

  • e-ISSN

    1748-0221

  • Volume of the periodical

    20

  • Issue of the periodical within the volume

    11

  • Country of publishing house

    GB - UNITED KINGDOM

  • Number of pages

    10

  • Pages from-to

    1-10

  • UT code for WoS article

    001615483400001

  • EID of the result in the Scopus database

    2-s2.0-105033813431