Characterization of a table-top “water window” soft X-ray source
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21460%2F17%3A00313993" target="_blank" >RIV/68407700:21460/17:00313993 - isvavai.cz</a>
Result on the web
—
DOI - Digital Object Identifier
—
Alternative languages
Result language
angličtina
Original language name
Characterization of a table-top “water window” soft X-ray source
Original language description
In this paper, a characterization of a table-top sized soft x-ray (SXR) source based on a Z-pinch capillary discharge in nitrogen and operating at He-like nitrogen wavelength of λ = 2.88 nm is presented. Measurements of pulse length and calculations of SXR photon flux have been carried out, the measurements were realized with respect to a nitrogen pressure in a range from 30 Pa to 350 Pa. Since the “water window” spectral region provides a natural contrast between oxygen and carbon absorption length, our SXR source emitting at λ = 2.88 nm is suitable for biological imaging.