THE EFFECT OF FRAME RATE AND CALIBRATION ON LUNG MONITORING WITH ELECTRICAL IMPEDANCE TOMOGRAPHY
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21460%2F20%3A00341351" target="_blank" >RIV/68407700:21460/20:00341351 - isvavai.cz</a>
Result on the web
<a href="https://doi.org/10.14311/CTJ.2019.4.01" target="_blank" >https://doi.org/10.14311/CTJ.2019.4.01</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.14311/CTJ.2019.4.01" target="_blank" >10.14311/CTJ.2019.4.01</a>
Alternative languages
Result language
angličtina
Original language name
THE EFFECT OF FRAME RATE AND CALIBRATION ON LUNG MONITORING WITH ELECTRICAL IMPEDANCE TOMOGRAPHY
Original language description
Electrical impedance tomography (EIT) is a non-invasive method that records changes in tissue impedance in the selected cross-section of the body and is mainly used to monitor patient’s lung ventilation. The aim of this study was to analyse the effect of the pre-set frame rate and how calibration of the EIT system affects final record when monitoring lungs using EIT system PulmoVista 500 (Dräger Medical, Germany). Ten healthy male volunteers were measured with frame rates 10, 20, 30, 40 and 50 Hz for analysis, followed by calibrated and uncalibrated measurements. The subjects were breathing spontaneously in a horizontal supine position. In case of the frame rate analysis, data were statistically analysed using ANOVA for repeated measures. Paired T-test was used for comparing data from calibrated and uncalibrated records. This study shows that the effect of both frame rate and calibration is negligible.
Czech name
—
Czech description
—
Classification
Type
J<sub>SC</sub> - Article in a specialist periodical, which is included in the SCOPUS database
CEP classification
—
OECD FORD branch
20601 - Medical engineering
Result continuities
Project
—
Continuities
S - Specificky vyzkum na vysokych skolach
Others
Publication year
2020
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Name of the periodical
Lékař a technika – Clinician and Technology
ISSN
0301-5491
e-ISSN
2336-5552
Volume of the periodical
49
Issue of the periodical within the volume
4
Country of publishing house
CZ - CZECH REPUBLIC
Number of pages
5
Pages from-to
107-111
UT code for WoS article
—
EID of the result in the Scopus database
2-s2.0-85091151064