Feasibility Evaluation of Metamaterial Microwave Sensors for Non-Invasive Blood Glucose Monitoring
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21460%2F21%3A00352684" target="_blank" >RIV/68407700:21460/21:00352684 - isvavai.cz</a>
Result on the web
<a href="https://doi.org/10.3390/s21206871" target="_blank" >https://doi.org/10.3390/s21206871</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.3390/s21206871" target="_blank" >10.3390/s21206871</a>
Alternative languages
Result language
angličtina
Original language name
Feasibility Evaluation of Metamaterial Microwave Sensors for Non-Invasive Blood Glucose Monitoring
Original language description
The use of microwave technology is currently under investigation for non-invasive estimation of glycemia in patients with diabetes. Due to their construction, metamaterial (MTM)-based sensors have the potential to provide higher sensitivity of the phase shift of the S-21 parameter (& ANGLE;S21) to changes in glucose concentration compared to standard microstrip transmission line (MSTL)-based sensors. In this study, a MSTL sensor and three MTM sensors with 5, 7, and 9 MTM unit cells are exposed to liquid phantoms with different dielectric properties mimicking a change in blood glucose concentration from 0 to 14 mmol/L. Numerical models were created for the individual experiments, and the calculated S-parameters show good agreement with experimental results, expressed by the maximum relative error of 8.89% and 0.96% at a frequency of 1.99 GHz for MSTL and MTM sensor with nine unit cells, respectively. MTM sensors with an increasing number of cells show higher sensitivity of 0.62 & DEG; per mmol/L and unit cell to blood glucose concentration as measured by changes in & ANGLE;S21. In accordance with the numerical simulations, the MTM sensor with nine unit cells showed the highest sensitivity of the sensors proposed by us, with an average of 3.66 & DEG; per mmol/L at a frequency of 1.99 GHz, compared to only 0.48 & DEG; per mmol/L for the MSTL sensor. The multi-cell MTM sensor has the potential to proceed with evaluation of human blood samples.</p>
Czech name
—
Czech description
—
Classification
Type
J<sub>imp</sub> - Article in a specialist periodical, which is included in the Web of Science database
CEP classification
—
OECD FORD branch
20601 - Medical engineering
Result continuities
Project
<a href="/en/project/LTC19031" target="_blank" >LTC19031: Development of metamaterial applicators for the regional hyperthermia system and evaluation of the accuracy of treatment planning algorithms</a><br>
Continuities
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)<br>S - Specificky vyzkum na vysokych skolach
Others
Publication year
2021
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Name of the periodical
Sensors
ISSN
1424-8220
e-ISSN
1424-8220
Volume of the periodical
21
Issue of the periodical within the volume
20
Country of publishing house
CH - SWITZERLAND
Number of pages
16
Pages from-to
1-16
UT code for WoS article
000713986300001
EID of the result in the Scopus database
2-s2.0-85116965801