Neutron Imaging with Micrometric Spatial Resolution
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21670%2F11%3A00191756" target="_blank" >RIV/68407700:21670/11:00191756 - isvavai.cz</a>
Result on the web
—
DOI - Digital Object Identifier
—
Alternative languages
Result language
angličtina
Original language name
Neutron Imaging with Micrometric Spatial Resolution
Original language description
The detection of neutrons with high spatial resolution by a pixelated silicon semiconductor silicon detector requires the use of a thin converter. A 6Li based converter was used for this purpose in our work. We demonstrate that unwanted gamma rays and electrons accompanying neutron beam can be effectively suppressed using pattern recognition of the analyzed particle tracks, where spectroscopic ability of the detector was advantageously used. Neutron detection with micrometric spatial resolution is reached by appropriate fitting of identified heavy charged particles by the Gaussian distribution.
Czech name
—
Czech description
—
Classification
Type
D - Article in proceedings
CEP classification
BG - Nuclear, atomic and molecular physics, accelerators
OECD FORD branch
—
Result continuities
Project
Result was created during the realization of more than one project. More information in the Projects tab.
Continuities
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Others
Publication year
2011
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Article name in the collection
IEEE Nuclear Science Symposium and Medical Imaging Conference 2011
ISBN
978-1-4673-0118-3
ISSN
1082-3654
e-ISSN
—
Number of pages
5
Pages from-to
1285-1289