BrachyView: Proof-of-principle of a novel in-body gamma camera for low dose-rate prostate brachytherapy
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21670%2F13%3A00216008" target="_blank" >RIV/68407700:21670/13:00216008 - isvavai.cz</a>
Result on the web
<a href="http://dx.doi.org/10.1118/1.4794487" target="_blank" >http://dx.doi.org/10.1118/1.4794487</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1118/1.4794487" target="_blank" >10.1118/1.4794487</a>
Alternative languages
Result language
angličtina
Original language name
BrachyView: Proof-of-principle of a novel in-body gamma camera for low dose-rate prostate brachytherapy
Original language description
BrachyView incorporates three Medipix2 detectors coupled to a multipinhole collimator. Three-dimensionally triangulated seed positions from multiple planar images are used to determine the seed placement in a PMMA prostate phantom in real time. MATLAB codes were used to test the reconstruction method and to optimize the device geometry.
Czech name
—
Czech description
—
Classification
Type
J<sub>x</sub> - Unclassified - Peer-reviewed scientific article (Jimp, Jsc and Jost)
CEP classification
JB - Sensors, detecting elements, measurement and regulation
OECD FORD branch
—
Result continuities
Project
<a href="/en/project/TA01010164" target="_blank" >TA01010164: Advanced Techniques for Detection of Ionizing Radiation</a><br>
Continuities
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Others
Publication year
2013
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Name of the periodical
Medical Physics
ISSN
0094-2405
e-ISSN
—
Volume of the periodical
40
Issue of the periodical within the volume
4
Country of publishing house
US - UNITED STATES
Number of pages
9
Pages from-to
—
UT code for WoS article
000317945900015
EID of the result in the Scopus database
—