BrachyView: Combining LDR seed positions with transrectal ultrasound imaging in a prostate gel phantom
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21670%2F17%3A00329986" target="_blank" >RIV/68407700:21670/17:00329986 - isvavai.cz</a>
Result on the web
<a href="http://dx.doi.org/10.1016/j.ejmp.2017.01.012" target="_blank" >http://dx.doi.org/10.1016/j.ejmp.2017.01.012</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1016/j.ejmp.2017.01.012" target="_blank" >10.1016/j.ejmp.2017.01.012</a>
Alternative languages
Result language
angličtina
Original language name
BrachyView: Combining LDR seed positions with transrectal ultrasound imaging in a prostate gel phantom
Original language description
Purpose: BrachyView is a novel in-body imaging system which aims to provide LDR brachytherapy seeds position reconstruction within the prostate in real-time. The first prototype is presented in this study: the probe consists of a gamma camera featuring three single cone pinhole collimators embedded in a tungsten tube, above three, high resolution pixelated detectors (Timepix). Methods: The prostate was imaged with a TRUS system using a sagittal crystal with a 2.5 mm slice thickness. Eleven needles containing a total of thirty 0.508 (UI)-I-125 seeds were implanted under ultrasound guidance. A CT scan was used to localise the seed positions, as well as provide a reference when performing the image co-registration between the BrachyView coordinate system and the TRUS coordinate system. An in-house visualisation software interface was developed to provide a quantitative 3D reconstructed prostate based on the TRUS images and co-registered with the LDR seeds in situ. A rigid body image registration was performed between the BrachyView and TRUS systems, with the BrachyView and CT-derived source locations compared. Results: The reconstructed seed positions determined by the BrachyView probe showed a maximum discrepancy of 1.78 mm, with 75% of the seeds reconstructed within 1 mm of their nominal location. An accurate co-registration between the BrachyView and TRUS coordinate system was established. Conclusions: The BrachyView system has shown its ability to reconstruct all implanted LDR seeds within a tissue equivalent prostate gel phantom, providing both anatomical and seed position information in a single interface.
Czech name
—
Czech description
—
Classification
Type
J<sub>imp</sub> - Article in a specialist periodical, which is included in the Web of Science database
CEP classification
—
OECD FORD branch
30224 - Radiology, nuclear medicine and medical imaging
Result continuities
Project
—
Continuities
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Others
Publication year
2017
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Name of the periodical
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS
ISSN
1120-1797
e-ISSN
1724-191X
Volume of the periodical
34
Issue of the periodical within the volume
February
Country of publishing house
GB - UNITED KINGDOM
Number of pages
10
Pages from-to
55-64
UT code for WoS article
000395371000007