Influence of various sterilization procedures on TiO2 nanotubes used for biomedical devices
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F70883521%3A28610%2F16%3A43874452" target="_blank" >RIV/70883521:28610/16:43874452 - isvavai.cz</a>
Result on the web
<a href="http://dx.doi.org/10.1016/j.bioelechem.2016.02.001" target="_blank" >http://dx.doi.org/10.1016/j.bioelechem.2016.02.001</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1016/j.bioelechem.2016.02.001" target="_blank" >10.1016/j.bioelechem.2016.02.001</a>
Alternative languages
Result language
angličtina
Original language name
Influence of various sterilization procedures on TiO2 nanotubes used for biomedical devices
Original language description
Sterilization is the final surface treatment procedure of all implantable devices and is one of the key factors which have to be considered before implementation. Since different sterilization procedures for all implantable devices influence mechanical properties as well as biological response, the influence of different sterilization techniques on titanium nanotubes was studied. Commonly used sterilization techniques such as autoclaving, ultra-violet light sterilization, hydrogen peroxide plasma sterilization as well as the not so frequently used gaseous oxygen plasma sterilization were used. Three different nanotube diameters; 15 nm, 50 nm and 100 nm were employed to study the effects of various sterilization techniques. It was observed that autoclave sterilization resulted in destruction of nanotubular features on all three studied nanotube diameters, while UV-light and both kinds of plasma sterilization did not cause any significant morphological changes on the surfaces. Differences between the sterilization techniques employed influenced cytocompatibility, especially in the case of nanotubes with 100 nm diameter.
Czech name
—
Czech description
—
Classification
Type
J<sub>x</sub> - Unclassified - Peer-reviewed scientific article (Jimp, Jsc and Jost)
CEP classification
JJ - Other materials
OECD FORD branch
—
Result continuities
Project
<a href="/en/project/LO1504" target="_blank" >LO1504: Centre of Polymer Systems Plus</a><br>
Continuities
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Others
Publication year
2016
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Name of the periodical
Bioelectrochemistry
ISSN
1567-5394
e-ISSN
—
Volume of the periodical
109
Issue of the periodical within the volume
Neuveden
Country of publishing house
CH - SWITZERLAND
Number of pages
8
Pages from-to
79-86
UT code for WoS article
000373418400010
EID of the result in the Scopus database
2-s2.0-84958568812