Experimental Measurements of Mechanical Properties of PUR Foam Used for Testing Medical Devices and Instruments Depending on Temperature, Density and Strain Rate

The result's identifiers

  • Result code in IS VaVaI

    <a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F71226401%3A_____%2F20%3AN0100403" target="_blank" >RIV/71226401:_____/20:N0100403 - isvavai.cz</a>

  • Alternative codes found

    RIV/68378297:_____/20:00541291 RIV/68407700:21220/20:00346579

  • Result on the web

    <a href="https://www.mdpi.com/1996-1944/13/20/4560" target="_blank" >https://www.mdpi.com/1996-1944/13/20/4560</a>

  • DOI - Digital Object Identifier

    <a href="http://dx.doi.org/10.3390/ma13204560" target="_blank" >10.3390/ma13204560</a>

Alternative languages

  • Result language

    angličtina

  • Original language name

    Experimental Measurements of Mechanical Properties of PUR Foam Used for Testing Medical Devices and Instruments Depending on Temperature, Density and Strain Rate

  • Original language description

    Rigid polyurethane (PUR) foam is products used as a biomedical material for medical device testing. Thermal stability is a very important parameter for evaluating the feasibility of use for testing surgical instrument load during drilling. This work aimed to perform experimental measurements to determine the dependence of the mechanical properties of a certified PUR on temperature, strain rate and density. Experimental measurements were realised for three types of the PUR samples with different density 10, 25 and 40 pounds per cubic foot. The samples were characterised in terms of their mechanical properties evaluated from tensile and compression tests at temperatures of 25 °C, 90 °C and 155 °C. Furthermore, the structures of the samples were characterised using optical microscope, their thermal properties were characterised by thermogravimetric analysis, and their density and stiffness with the effect of temperature was monitored. The results show that it is optimal not only for mechanical testing but also for testing surgical instruments that generate heat during machining. On the basis of experimental measurements and evaluations of the obtained values, the tested materials are suitable for mechanical testing of medical devices. At the same time, this material is also suitable for testing surgical instruments that generate heat during machining.

  • Czech name

  • Czech description

Classification

  • Type

    J<sub>imp</sub> - Article in a specialist periodical, which is included in the Web of Science database

  • CEP classification

  • OECD FORD branch

    20301 - Mechanical engineering

Result continuities

  • Project

    <a href="/en/project/GA17-25821S" target="_blank" >GA17-25821S: Evaluation and modelling of thermal field during machining of non-metallic materials</a><br>

  • Continuities

    P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)

Others

  • Publication year

    2020

  • Confidentiality

    S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů

Data specific for result type

  • Name of the periodical

    Materials

  • ISSN

    1996-1944

  • e-ISSN

    1996-1944

  • Volume of the periodical

    13

  • Issue of the periodical within the volume

    20

  • Country of publishing house

    CH - SWITZERLAND

  • Number of pages

    13

  • Pages from-to

    4560

  • UT code for WoS article

    000585574600001

  • EID of the result in the Scopus database

    2-s2.0-85093921816