Long-term outcomes of the pentaspline pulsed-field ablation catheter for the treatment of paroxysmal atrial fibrillation: results of the prospective, multicentre FARA-Freedom Study

The result's identifiers

  • Result code in IS VaVaI

    <a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00023001%3A_____%2F24%3A00084688" target="_blank" >RIV/00023001:_____/24:00084688 - isvavai.cz</a>

  • Alternative codes found

    RIV/65269705:_____/24:00079833 RIV/00023884:_____/24:00010017

  • Result on the web

    <a href="https://academic.oup.com/europace/article/26/3/euae053/7612551?login=true#445702905" target="_blank" >https://academic.oup.com/europace/article/26/3/euae053/7612551?login=true#445702905</a>

  • DOI - Digital Object Identifier

    <a href="http://dx.doi.org/10.1093/europace/euae053" target="_blank" >10.1093/europace/euae053</a>

Alternative languages

  • Result language

    angličtina

  • Original language name

    Long-term outcomes of the pentaspline pulsed-field ablation catheter for the treatment of paroxysmal atrial fibrillation: results of the prospective, multicentre FARA-Freedom Study

  • Original language description

    AimsPulmonary vein isolation (PVI) is a well-established strategy for the treatment of paroxysmal atrial fibrillation (PAF). Despite randomized controlled trials and real-world data showing the promise of pulsed-field ablation (PFA) for this treatment, long-term efficacy and safety data demonstrating single-procedure outcomes off antiarrhythmic drugs remain limited. The aim of the FARA-Freedom Study was to evaluate the long-term efficacy and safety of PFA using the pentaspline catheter for PAF.Methods and resultsFARA-Freedom, a prospective, non-randomized, multicentre study, enrolled patients with PAF undergoing de novo PVI with PFA, who were followed for 12 months with weekly transtelephonic monitoring and a 72-h Holter ECG at 6 and 12 months. The primary safety endpoint was a composite of device- or procedure-related serious adverse events out to 7 days post-ablation and PV stenosis or atrioesophageal (AE) fistula out to 12 months. Treatment success is a composite of acute PVI and chronic success, which includes freedom from any documented atrial tachyarrhythmia longer than 30 s, use of antiarrhythmic drugs or cardioversion after a 3-month blanking period, or use of amiodarone or repeat ablation at any time. The study enrolled 179 PAF patients (62 ± 10 years, 39% female) at 13 centres. At the index procedure, all PVs were successfully isolated with the pentaspline PFA catheter. Procedure and left atrial dwell times, with a 20-min waiting period, were 71.9 ± 17.6 and 41.0 ± 13.3 min, respectively. Fluoroscopy time was 11.5 ± 7.4 min. Notably, monitoring compliance was high, with 88.4 and 90.3% with weekly events and 72-h Holter monitors, respectively. Freedom from the composite primary effectiveness endpoint was 66.6%, and 41 patients had atrial tachyarrhythmia recurrence, mostly recurrent atrial fibrillation (31 patients). The composite safety endpoint occurred in two patients (1.1%), one tamponade and one transient ischaemic attack. There was no coronary spasm, PV stenosis, or AE fistula. There were four cases of transient phrenic nerve palsy, but all resolved during the index procedure.ConclusionIn this prospective, non-randomized, multicentre study, PVI using a pentaspline PFA catheter was effective in treating PAF patients despite rigourous endpoint definitions and high monitoring compliance and demonstrated favourable safety.

  • Czech name

  • Czech description

Classification

  • Type

    J<sub>imp</sub> - Article in a specialist periodical, which is included in the Web of Science database

  • CEP classification

  • OECD FORD branch

    30201 - Cardiac and Cardiovascular systems

Result continuities

  • Project

  • Continuities

    N - Vyzkumna aktivita podporovana z neverejnych zdroju

Others

  • Publication year

    2024

  • Confidentiality

    S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů

Data specific for result type

  • Name of the periodical

    Europace

  • ISSN

    1099-5129

  • e-ISSN

    1532-2092

  • Volume of the periodical

    26

  • Issue of the periodical within the volume

    3

  • Country of publishing house

    GB - UNITED KINGDOM

  • Number of pages

    7

  • Pages from-to

    "art. no. euae053"

  • UT code for WoS article

    001183455000001

  • EID of the result in the Scopus database

    2-s2.0-85187712076