Two-year experience with the commercial Gamma Knife Check software
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00023884%3A_____%2F16%3A%230006777" target="_blank" >RIV/00023884:_____/16:#0006777 - isvavai.cz</a>
Result on the web
—
DOI - Digital Object Identifier
—
Alternative languages
Result language
angličtina
Original language name
Two-year experience with the commercial Gamma Knife Check software
Original language description
The Gamma Knife Check software is an FDA approved second check system for dose calculations in Gamma Knife radiosurgery. The purpose of this study was to evaluate the accuracy and the stability of the commercial software package as a tool for independent dose verification. The Gamma Knife Check software version 8.4 was commissioned for a Leksell Gamma Knife Perfexion and a 4C unit at the University of Pittsburgh Medical Center in May 2012. Independent dose verifications were performed using this software for 319 radiosurgery cases on the Perfexion and 283 radiosurgery cases on the 4C units. The cases on each machine were divided into groups according to their diagnoses, and an averaged absolute percent dose difference for each group was calculated. The percentage dose difference for each treatment target was obtained as the relative difference between the Gamma Knife Check dose and the dose from the tissue maximum ratio algorithm (TMR 10) from the GammaPlan software version 10 at the reference point. For treatment plans with imaging skull definition, results obtained from the Gamma Knife Check software using the measurement-based skull definition method are used for comparison. The collected dose difference data were also analyzed in terms of the distance from the treatment target to the skull, the number of treatment shots used for the target, and the gamma angles of the treatment shots. The averaged percent dose differences between the Gamma Knife Check software and the GammaPlan treatment planning system are 0.3%, 0.89%, 1.24%, 1.09%, 0.83%, 0.55%, 0.33%, and 1.49% for the trigeminal neuralgia, acoustic neuroma, arteriovenous malformation (AVM), meningioma, pituitary adenoma, glioma, functional disorders, and metastasis cases on the Perfexion unit. The corresponding averaged percent dose differences for the 4C unit are 0.33%, 1.2%, 2.78% 1.99%, 1.4%, 1.92%, 0.62%, and 1.51%, respectively. The dose difference is, i
Czech name
—
Czech description
—
Classification
Type
J<sub>x</sub> - Unclassified - Peer-reviewed scientific article (Jimp, Jsc and Jost)
CEP classification
FH - Neurology, neuro-surgery, nuero-sciences
OECD FORD branch
—
Result continuities
Project
—
Continuities
N - Vyzkumna aktivita podporovana z neverejnych zdroju
Others
Publication year
2016
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Name of the periodical
Journal of Applied Clinical Medical Physics
ISSN
1526-9914
e-ISSN
—
Volume of the periodical
17
Issue of the periodical within the volume
4
Country of publishing house
US - UNITED STATES
Number of pages
11
Pages from-to
95-105
UT code for WoS article
000380374900010
EID of the result in the Scopus database
2-s2.0-84995459590