Fetal Dose Evaluation for Pregnant Patients on Leksell Gamma Knife Perfexion/Icon
The result's identifiers
Result code in IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00023884%3A_____%2F24%3A00009624" target="_blank" >RIV/00023884:_____/24:00009624 - isvavai.cz</a>
Result on the web
<a href="https://karger.com/sfn/article-abstract/102/2/65/893970/Fetal-Dose-Evaluation-for-Pregnant-Patients-on?redirectedFrom=fulltext" target="_blank" >https://karger.com/sfn/article-abstract/102/2/65/893970/Fetal-Dose-Evaluation-for-Pregnant-Patients-on?redirectedFrom=fulltext</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1159/000535366" target="_blank" >10.1159/000535366</a>
Alternative languages
Result language
angličtina
Original language name
Fetal Dose Evaluation for Pregnant Patients on Leksell Gamma Knife Perfexion/Icon
Original language description
It is a normal procedure to avoid the application of ionizing radiation during pregnancy. In very rare occasions treatment can be performed but doses to the fetus must be evaluated and reported and the patient must sign informed consent. There can occur two types of damage caused by ionizing radiation – deterministic and stochastic effects. Deterministic effects may occur after reaching a certain threshold (100 mGy for this study), meanwhile, stochastic effects have no limit and their probability rises with dose. This study focuses on deterministic effects. This study compares the dose measured on phantom for the area of the pelvis and the dose measured on 3 patients with dosimeters positioned on the pelvis irradiated on Leksell Gamma Knife Perfexion/Icon. The mean dose for measurement on phantom for pelvis was 0.73 ± 0.76 mGy and for patients, it was 1.28 mGy, 0.493 mGy, 0.549 mGy which is 80 times lower, 200 times lower, 180 times lower than the threshold for deterministic effects, respectively. The measurement carried on phantom served as the base for drafting informed consent and provided initial proof that treatment can be safely delivered. Measurements performed on patients only confirmed that irradiation of pregnant patients on Leksell Gamma Knife Perfexion/Icon is safe relative to the deterministic effects. Nevertheless, pregnant patients should be treated with ionizing radiation only in very extraordinary situations.
Czech name
—
Czech description
—
Classification
Type
J<sub>imp</sub> - Article in a specialist periodical, which is included in the Web of Science database
CEP classification
—
OECD FORD branch
30103 - Neurosciences (including psychophysiology)
Result continuities
Project
—
Continuities
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Others
Publication year
2024
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Name of the periodical
Stereotactic and Functional Neurosurgery
ISSN
1011-6125
e-ISSN
—
Volume of the periodical
102
Issue of the periodical within the volume
2
Country of publishing house
DE - GERMANY
Number of pages
9
Pages from-to
65-73
UT code for WoS article
001164053200001
EID of the result in the Scopus database
2-s2.0-85182552454