Original language description
Background: The laryngeal mask airway ( LMA ) is a well-established device for securing the airway, typically inserted using a blind technique. However, blind insertion may result in suboptimal alignment and the need for reinsertion, which can increase the risk of adverse events. Therefore, in this pilot study, we aimed to conduct a preliminary evaluation and gather initial clinical experience with the new video laryngeal mask, SaCoVLM (TM). Methods and materials: In this prospective observational study, the SaCoVLM (TM) was used for airway management in patients undergoing general anaesthesia. We assessed key quality parameters commonly used to compare the performance of different laryngeal masks. Primary outcome was first attempt success rate. Secondary outcomes were time to achieve adequate ventilation, oropharyngeal leak pressure (OLP), and the incidence of adverse events. Results: A total of 50 patients undergoing general anaesthesia were enrolled. The first-attempt insertion success rate was 96%. The mean time to achieve adequate ventilation was 13 +/- 2.4 s, and the mean OLP was 38 +/- 3.8 mbar. The overall incidence of adverse events related to video laryngeal mask usage was 6%, with no serious adverse events observed. Conclusion: The SaCoVLM (TM) appears to be a promising alternative for airway management in patients undergoing general anaesthesia, demonstrating favourable performance across key clinical parameters.
Czech name
—
Czech description
—
Classification
Type
J<sub>imp</sub> - Article in a specialist periodical, which is included in the Web of Science database
CEP classification
—
OECD FORD branch
30223 - Anaesthesiology
Result continuities
Project
—
Continuities
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Others
Publication year
2026
Confidentiality
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Data specific for result type
Name of the periodical
BMC ANESTHESIOLOGY
ISSN
1471-2253
e-ISSN
—
Volume of the periodical
26
Issue of the periodical within the volume
1
Country of publishing house
GB - UNITED KINGDOM
Number of pages
7
Pages from-to
-
UT code for WoS article
001696791900002
EID of the result in the Scopus database
—