Development of a new primary humidity measurement standard

The result's identifiers

  • Result code in IS VaVaI

    <a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00177016%3A_____%2F19%3AN0000015" target="_blank" >RIV/00177016:_____/19:N0000015 - isvavai.cz</a>

  • Alternative codes found

    RIV/00177016:_____/20:N0000016

  • Result on the web

    <a href="https://iopscience.iop.org/article/10.1088/1361-6501/ab4a54" target="_blank" >https://iopscience.iop.org/article/10.1088/1361-6501/ab4a54</a>

  • DOI - Digital Object Identifier

    <a href="http://dx.doi.org/10.1088/1361-6501" target="_blank" >10.1088/1361-6501</a>

Alternative languages

  • Result language

    angličtina

  • Original language name

    Development of a new primary humidity measurement standard

  • Original language description

    A new primary humidity generator was developed at the National Standards Authority of Ireland (NSAI). The principles of operation and overall design of the generator are discussed. The flow regime within the saturators was analysed using computational fluid dynamics (CFD) to determine if sufficient mixing takes place within the final saturator. This was later investigated by experiment and it was found that sufficient mixing takes place for flow rates up to 3 l min−1 at standard flow conditions of 101 325 Pa and 293.15 K. The uncertainty due to non-ideal saturation at 90 °C dp is discussed. This uncertainty was analysed with respect to the effect that the flow speed and latent heating in the saturators has on the saturation efficiency, separately. The uncertainty associated with saturation efficiency at 90 °C dew point was estimated from the temperature gradients along the saturation path arising due to latent heating effects. The standard uncertainty due to non-ideal saturation was found to be  ±3 mK dp.

  • Czech name

  • Czech description

Classification

  • Type

    J<sub>imp</sub> - Article in a specialist periodical, which is included in the Web of Science database

  • CEP classification

  • OECD FORD branch

    20205 - Automation and control systems

Result continuities

  • Project

    <a href="/en/project/8B16006" target="_blank" >8B16006: Expansion of European research capabilities in humidity measurement</a><br>

  • Continuities

    P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)

Others

  • Publication year

    2019

  • Confidentiality

    S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů

Data specific for result type

  • Name of the periodical

    Measurement Science and Technology

  • ISSN

    09570233

  • e-ISSN

    13616501

  • Volume of the periodical

    2020

  • Issue of the periodical within the volume

    31

  • Country of publishing house

    GB - UNITED KINGDOM

  • Number of pages

    10

  • Pages from-to

    1-10

  • UT code for WoS article

    000499383000001